G accuracy (when Dx ≤ 0.1)
Equation 5.
Bx + Da (S)
1
Bx = 2πfCx =——————
2πfLx
Dx
Measured D value
Bx
Measured B value (S)
Da
Absolute accuracy of D
f
Test frequency (Hz)
Cx
Measured C value (F)
Lx
Measured L value (H)
where the accuracy of G is applied to Cp-G measurements.
Absolute accuracy of Rp (when Dx ≤ 0.1)
Equation 6.
Rpx × Da
± —————————(Ω)
Dx± Da
Rpx Measured Rp value (Ω)
Dx
Measured D value
Da
Absolute accuracy of D
Absolute accuracy of Rs (when Dx ≤ 0.1)
Equation 7. Xx × Da(Ω)
1
Xx =——————= 2πfLx
2πfCx
Dx
Measured D value
Xx
Measured X value (Ω)
Da
Absolute accuracy of D
f
Test frequency (Hz)
Cx
Measured C value (F)