E4980A-301 Datasheet, PDF (8/38 Pages) Keysight Technologies – Precision LCR Meter | |||
|
◁ |
08 | Keysight | E4980A Precision LCR Meter 20 Hz to 2 MHz, E4980AL Precision LCR Meter 20 Hz to 300 kHz/500 kHz/1 MHz â Data Sheet
G accuracy (when Dx ⤠0.1)
Equation 5.
Bx + Da (S)
1
Bx = 2ÏfCx =ââââââ
2ÏfLx
Dx
Measured D value
Bx
Measured B value (S)
Da
Absolute accuracy of D
f
Test frequency (Hz)
Cx
Measured C value (F)
Lx
Measured L value (H)
where the accuracy of G is applied to Cp-G measurements.
Absolute accuracy of Rp (when Dx ⤠0.1)
Equation 6.
Rpx à Da
± âââââââââ(â¦)
Dx± Da
Rpx Measured Rp value (Ω)
Dx
Measured D value
Da
Absolute accuracy of D
Absolute accuracy of Rs (when Dx ⤠0.1)
Equation 7. Xx à Da(â¦)
1
Xx =ââââââ= 2ÏfLx
2ÏfCx
Dx
Measured D value
Xx
Measured X value (Ω)
Da
Absolute accuracy of D
f
Test frequency (Hz)
Cx
Measured C value (F)
Lx
Measured L value (H)
|
▷ |