English
Language : 

E4980A-301 Datasheet, PDF (25/38 Pages) Keysight Technologies – Precision LCR Meter
25 | Keysight | E4980A Precision LCR Meter 20 Hz to 2 MHz, E4980AL Precision LCR Meter 20 Hz to 300 kHz/500 kHz/1 MHz – Data Sheet
Effect of cable length (Short offset)
Table 46. The following value is added to Rs when the cable is extended.
Cable length
1m
2m
4m
0.25 mΩ
0.5 mΩ
1 mΩ
Temperature coefficient Kt
Table 47. Temperature coefficient Kt is given below.
Temperature [°C]
Kt
0 - 18
4
18 - 28
1
28 - 55
4