English
Language : 

E4980A-301 Datasheet, PDF (32/38 Pages) Keysight Technologies – Precision LCR Meter
32 | Keysight | E4980A Precision LCR Meter 20 Hz to 2 MHz, E4980AL Precision LCR Meter 20 Hz to 300 kHz/500 kHz/1 MHz – Data Sheet
Table 55. E4980A-005 measurement time [ms] (DC bias: OFF)
Measurement
time mode
Test frequency
20 Hz 100 Hz 1 kHz
10 kHz 100 kHz 1 MHz
1 LONG
1190 650
590
580
570
570
2 MED
1150 380
200
180
180
180
3 SHORT
1040 240
37
25
23
23
2 MHz
570
180
23
10
1
1. LONG
2. MED
3. SHORT
0.1
0.01
0.001
20
100
1k
10k
100k
Test frequency [Hz]
Figure 10. Measurement time (DC bias: OFF, E4980A-005)
1M 2M
Table 56. E4980AL measurement time [ms]
Measurement
time mode
Test frequency
20 Hz 100 Hz
1 LONG
729
423
2 MED
650
250
3 SHORT
579
149
1 kHz
363
140
26
10
10 kHz
353
122
14
100 kHz
343
119
12
1 MHz
343
118
12
1
1. LONG
2. MED
3. SHORT
0.1
0.01
0.001
20
100
1k
10k
100k
Test frequency [Hz]
Figure 11. Measurement time (E4980AL)
1M 2M