English
Language : 

E4980A-301 Datasheet, PDF (6/38 Pages) Keysight Technologies – Precision LCR Meter
06 | Keysight | E4980A Precision LCR Meter 20 Hz to 2 MHz, E4980AL Precision LCR Meter 20 Hz to 300 kHz/500 kHz/1 MHz – Data Sheet
Measurement display ranges
Table 10 shows the range of measured value that can be displayed on the screen.
For the effective measurement ranges, refer to Figure 1 impedance measurement
accuracy example.
Table 10. Allowable display ranges for measured values
Parameter
Measurement display range
Cs, Cp
± 1.000000 aF to 999.9999 EF
Ls, Lp
± 1.000000 aH to 999.9999 EH
D
± 0.000001 to 9.999999
Q
± 0.01 to 99999.99
R, Rs, Rp, X, Z, Rdc
± 1.000000 aΩ to 999.9999 EΩ
G, B, Y
± 1.000000 aS to 999.9999 ES
Vdc
± 1.000000 aV to 999.9999 EV
Idc
± 1.000000 aA to 999.9999 EA
θr
± 1.000000 arad to 3.141593 rad
θd
± 0.0001 deg to 180.0000 deg
Δ%
± 0.0001 % to 999.9999 %
a: 1 x 10-18, E: 1 x 1018