Performance characteristics
Rise/fall time (ms)1
For resistive measurement2
±1 μA
±10 μA
±100 μA
±1 mA
±10 mA
±120 mA
Remote sense operating range
Temperature coefficient
Guard output resistance
Noise 10 Hz to 20 MHz (peak-peak)
Output voltage overshoot, ±(% of output + offset)2
Programming language
Maximum sense lead resistance
Voltage line regulation
Voltage load regulation
Current line regulation
Current load regulation
Recommended calibration interval
U2722A
170.0
18.0
6.0
1.0
1.0
1.0
U2723A
15.0
5.0
1.0
1.0
1.0
1.0
Ensure that the maximum voltage between the OUTPUT+ and SENSE+,
OUTPUT–, and SENSE– does not exceed 3 V.
Maximum change in output/readback per °C after a 30-minute
warm-up is 0.15.
0.2 kΩ
100 mV typical into a resistive load (floating mode).
During turn-on or turn-off, the output plus overshoot < 0.1% + 10 mV.
SCPI (Standard Commands for Programmable Instruments)
1 MΩ for rated accuracy
0.01% of range
0.01% + 100 μV
0.04% of range
0.04% + 100 μA
One year
NOTE
–– All channels are isolated from the ground and from each other. Isolation is +60 VDC, Category 1.
–– All specifications are based on three hours warm-up time.
–– The measurement accuracy value is x ( 1 + a * y ), where,
–– x = accuracy specification at room temperature,
–– a = temperature coefficient, and
–– y = temperature change from room temperature in °C
Noise 10 Hz to 20 MHz (Peak-peak)
Voltage range
2V
20 V
1 μA
50 mV
50 mV
10 μA
50 mV
50 mV
Current range
100 μA
1 mA
50 mV
50 mV
50 mV
50 mV
10 mA
30 mV
30 mV
120 mA
30 mV
30 mV
1 Drive 50% of 1 V or 10 V output with a resistive load. Rise time is from 10% to 90% of program voltage change at maximum current. Fall time is from
90% to 10% of program voltage change at maximum current.
2 Measurements obtained are per default bandwidth setting.
57