U2781A Datasheet, PDF (42/76 Pages) Keysight Technologies – USB Modular Products | |||
|
◁ |
System accuracy specifications
The U2802A system accuracy specifications are shown in Table 1, Table 2, and Table 3. These measurements are derived
from the U2802A and DAQ input accuracy specifications without including the thermocouple error. Refer to âCalculating
System Accuracyâ section in the Keysight U2802A 31-Channel Thermocouple Input Device Userâs Guide for calculation
methodology.
Table 1. Measurement accuracy of the U2355A and U2356A at 23 °C ± 5 °C
Thermocouple measurement accuracy (U2355A, U2356A @ 23 °C ± 5 °C)
ITS-90 Temperature range (°C) Optimum measurement range (°C)
T/C type
Low
High
Low
High
Without
averaging
(± °C)
50 points
averaging
(± °C)
B
0
1820
1100
1820
1.9
1.2
400
1100
4.4
2.5
E
â270
1000
â150
1000
1.7
1.6
â200
â150
2.4
2.3
J
â210
1200
â150
1200
1.6
1.5
â210
â150
2.7
2.6
K
â270
1372
â100
1200
1.5
1.4
â200
â100
2.7
2.6
N
â270
1300
â100
1300
1.5
1.3
â200
â100
3.0
2.7
R
â50
1768
300
1760
2.0
1.4
â50
300
5.0
3.1
S
â50
1768
400
1760
2.1
1.6
â50
400
4.5
2.8
T
â270
400
â100
400
1.5
1.4
â200
â100
2.7
2.5
500 points
averaging
(± °C)
1.0
2.0
1.6
2.3
1.5
2.5
1.4
2.6
1.3
2.6
1.3
2.6
1.4
2.4
1.4
2.5
Table 2. Measurement accuracy of the U2355A at 0 to 18 °C and 28 to 45 °C
Thermocouple measurement accuracy (U2355A @ 0 to 18 °C and 28 to 45 °C)
T/C type ITS-90 Temperature range (°C) Optimum measurement range (°C)
Low
High
Low
High
Without
averaging
(± °C)
50 points
averaging
(± °C)
B
0
1820
1100
1820
3.4
2.4
400
1100
7.5
3.6
E
â270
1000
â150
1000
2.7
2.6
â200
â150
3.8
3.6
J
â210
1200
â150
1200
2.5
2.4
â210
â150
4.2
4.0
K
â270
1372
â100
1200
2.9
2.8
â200
â100
4.3
4.0
N
â270
1300
â100
1300
2.6
2.5
â200
â100
4.9
4.2
R
â50
1768
300
1760
3.8
3.1
â50
300
8.5
4.6
S
â50
1768
400
1760
4.2
3.4
â50
400
7.7
4.2
T
â270
400
â100
400
2.4
2.2
â200
â100
4.3
4.3
500 points
averaging
(± °C)
2.2
2.2
2.5
3.6
2.4
3.9
2.8
3.9
2.5
4.0
3.0
3.3
3.2
3.1
2.2
3.9
42
|
▷ |