English
Language : 

U2781A Datasheet, PDF (23/76 Pages) Keysight Technologies – USB Modular Products
DC Characteristics
Accuracy Specifications
Model Number
Unipolar Range (V)
10
5
2.5
1.25
Bipolar Range (V)
10
5
2.5
1.25
U2351A, U2352A, U2353A, U2354A
Analog Input
Offset Error (mV)1
Gain Error (mV)
1.5
2.0
1.5
2.0
1.0
1.0
1.0
1.0
Offset Error (mV)1
Gain Error (mV)
1.0
2.0
1.0
2.0
1.0
1.5
1.0
1.5
Accuracy (% of reading + offset error)2
0.04% + 1.5 mV
0.08% + 1.5 mV
0.08% + 1.0 mV
0.16% + 1.0 mV
Accuracy (% of reading + offset error)2
0.02% + 1.0 mV
0.04% + 1.0 mV
0.06% + 1.0 mV
0.12% + 1.0 mV
Model Number
Unipolar Range (V)
10
5
2.5
1.25
Bipolar Range (V)
10
5
2.5
1.25
Offset Error (mV)1
1.0
1.0
1.0
1.0
Offset Error (mV)1
1.0
1.0
1.0
1.0
U2355A, U2356A
Gain Error (mV)
1.5
1.5
1.0
1.0
Gain Error (mV)
2.0
2.0
1.5
1.5
Accuracy (% of reading + offset error)2
0.03% + 1.0 mV
0.06% + 1.0 mV
0.08% + 1.0 mV
0.16% + 1.0 mV
Accuracy (% of reading + offset error)2
0.02% + 1.0 mV
0.04% + 1.0 mV
0.06% + 1.0 mV
0.12% + 1.0 mV
Model
Unipolar Range (V)
10
5
4
2.5
2
1
0.5
0.4
0.1
Offset Error (mV)1
1.5
1.5
1.5
1.0
1.0
1.0
1.0
1.0
1.0
U2331A
Gain Error (mV)
4.0
2.0
2.0
1.5
1.0
1.0
1.0
1.0
1.0
–– The above specifications are typical for 23 °C.
–– Specifications are for 20 minutes warm-up and self calibration.
–– The measurement are calculated with 100 points avearaging of data.
Accuracy (% of reading + offset error)2
0.08% + 1.5 mV
0.08% + 1.5 mV
0.10% + 1.5 mV
0.12% + 1.0 mV
0.10% + 1.0 mV
0.20% + 1.0 mV
0.41% + 1.0 mV
0.51% + 1.0 mV
2.04% + 1.0 mV
1 Offset error is measured at midscale of full range.
2 Accuracy = ± [% of |(Gain error / (Measured value – Midscale of FSR))| + Offset error]
23