English
Language : 

U2044XA Datasheet, PDF (5/25 Pages) Keysight Technologies – U2040 X-Series Wide Dynamic Range Power Sensors
05 | Keysight | U2040 X-Series Wide Dynamic Range Power Sensors – Data Sheet
20 automatic pulse parameter measurements
The U2040 X-Series offers simultaneous pulse parameter characterization of up to
20 pulses within a single capture. Individual pulse duration, period, duty cycle, rise time,
fall time and other pulse parameters can be queried through the following SCPI codes:
TRACe:MEASurement:PULSe[1-20], and TRACe:MEASurement:TRANsition[1-20].
Together with a system’s rise time and fall time of 100 ns and video bandwidth of
5 MHz, the U2040 X-Series enables a minimum measurable pulse width of 250 ns with
its sampling interval of 50 ns. Users can quickly and accurately measure the output power
and pulse parameters of pulses for radar pulse component design or manufacturing.
1
2
3
4
5
6
........
19
20
Pulse parameter
Duty cycle
Pulse duration
Pulse period
Pulse separation
Negative transition duration (fall time)
Occurrence of a negative transition relative to trigger instant
Positive transition duration (rise time)
Occurrence of a positive transition relative to trigger instant
SCPI command
TRAC:MEAS:PULS[1-20]:DCYC?
TRAC:MEAS:PULS[1-20]:DUR?
TRAC:MEAS:PULS[1-20]:PER?
TRAC:MEAS:PULS[1-20]:SEP?
TRAC:MEAS:TRAN[1-20]:NEG:DUR?
TRAC:MEAS:TRAN[1-20]:NEG:OCC?
TRAC:MEAS:TRAN[1-20]:POS:DUR?
TRAC:MEAS:TRAN[1-20]:POS:OCC?
Figure 3. The U2040 X-Series offers simultaneous analysis of up to 20 pulses within a single capture