English
Language : 

U2044XA Datasheet, PDF (11/25 Pages) Keysight Technologies – U2040 X-Series Wide Dynamic Range Power Sensors
11 | Keysight | U2040 X-Series Wide Dynamic Range Power Sensors – Data Sheet
Maximum SWR
Frequency band
Power Level
10 MHz to 6 GHz
> 6 GHz to 18 GHz
U2041XA/42XA
-70 to +15 dBm
+15 to +26 dBm
< 1.2
< 1.29
U2043XA/44XA
-70 to +15 dBm
+15 to +26 dBm
< 1.20
< 1.29
< 1.26
< 1.30
Frequency band
Power Level
10 MHz to 30 MHz
> 30 MHz to 50 MHz
> 50 MHz to 100 MHz
> 100 MHz to 11.5 GHz
> 11.5 GHz to 30 GHz
> 30 GHz to 33 GHz
U2049XA
-70 to +15 dBm
+15 to +20 dBm
< 2.18
< 2.21
< 1.35
< 1.37
< 1.22
< 1.24
< 1.17
< 1.29
< 1.33
< 1.21
< 1.33
< 1.36
Calibration uncertainty
Definition: Uncertainty resulting from non-linearity in the U2040 X-Series detection and correction process. This can be considered as
a combination of traditional linearity, calibration factor and temperature specifications and the uncertainty associated with the internal
calibration process.
Frequency band
10 MHz to 30 MHz
> 30MHz to 500 MHz
> 500 MHz to 1 GHz
> 1 GHz to 6 GHz
> 6 GHz to 10 GHz
> 10 GHz to 18 GHz
> 18 GHz to 26.5 GHz
> 26.5 GHz to 33 GHz
U2041XA/42XA
4.4%
3.7%
3.7%
3.7%
-
-
-
-
Average mode
U2043XA/44XA
4.4%
3.7%
3.7%
3.7%
3.7%
4.0%
-
-
U2049XA
4.5%
3.9%
3.8%
3.9%
4.0%
4.2%
4.9%
5.6%
Frequency band
10 MHz to 30 MHz
> 30MHz to 500 MHz
> 500 MHz to 1 GHz
> 1 GHz to 6 GHz
> 6 GHz to 10 GHz
> 10 GHz to 18 GHz
> 18 GHz to 26.5 GHz
> 26.5 GHz to 33 GHz
VBW OFF/HIGH 1
U2042XA
U2044XA
5.7%
5.7%
5.2%
5.2%
5.2%
5.2%
5.3%
5.3%
-
5.3%
-
5.4%
-
-
-
-
Normal mode
U2049XA
4.5%
4.1%
3.9%
4.0%
4.1%
4.3%
5.0%
5.7%
U2042XA
4.4%
3.7%
3.7%
3.7%
-
-
-
-
VBW MED/LOW
U2044XA
4.4%
3.7%
3.7%
3.7%
3.7%
4.0%
-
-
U2049XA
4.5%
3.9%
3.9%
4.0%
4.1%
4.2%
4.9%
5.6%
1.  Specification valid for environment up to 70% relative humidity. Additional 1.6% to be included for
environment up to 95% relative humidity.