33621A Datasheet, PDF (23/24 Pages) Keysight Technologies – Trueform Waveform Generators
23 | Keysight | M33600A Series Trueform Waveform Generators - Data Sheet
Deinitions
Speciication (spec)
The warranted performance of a calibrated instrument that has been stored for a
minimum of 2 hours within the operating temperature range of 0 to 55 ºC and after
a 1-hour warm up period. All speciications account for the effects of measurement
and calibration-source uncertainties, and were created in compliance with
ISO-17025 methods. Data published in this document are speciications (spec)
only where speciically indicated.
Typical (typ)
The characteristic performance that 80% or more of manufactured instruments will
meet. This data is not warranted, does not include measurement or calibration-source
uncertainty, and is valid only at room temperature (approximately 23 ºC).
Nominal (nom)
The mean or average characteristic performance, or the value of an attribute that is
determined by design such as a connector type, physical dimension, or operating speed.
This data is not warranted and is measured at room temperature (approximately 23 ºC).
Measured (meas)
An attribute measured during product development for the purpose of communicating
expected performance. This data is not warranted and is measured at room temperature
(approximately 23 ºC).