E1987A Datasheet, PDF (18/27 Pages) Keysight Technologies – cdma2000/IS-95/AMPS
18 | Keysight | E1962B cdma2000/IS-95/AMPS Mobile Test Application - Technical Overview
CDMA RF analyzer (measurements only)
Frequency range (reverse channels)
Band
Frequency
Channel
US cellular band
815 to 849 MHz
1 to 799
991 to 1023
1024 to 1323
1324 to 1424
US PCS band
1850 to 1910 MHz 1 to 1199
Korean PCS band
1750 to 1780 MHz 0 to 599
Japan CDMA band
887 to 925 MHz
1 to 799
801 to 1039
1041 to 1199
1201 to 1600
IMT-2000 band
1920 to 1980 MHz 0 to 1199
NMT-450 band
410 to 484 MHz
1 to 300
539 to 871
1039 to 1473
1792 to 2016
Secondary 800 band
806 to 824
896 to 901 MHz
0 to 719
720 to 919
US PCS 1.9 GHz band 1850 to 1915 MHz 0 to 1299
AWS band
1710 to 1755 MHz 0 to 899
Cellular upper 700 band
0 to 240
400 MHz European
PAMR
410 to 420
450 to 460
479 to 483 MHz
1 to 2016
800 MHz PAMR band 870 to 876 MHz
0 to 239
700 MHz public safety 787 to 799 MHz
band
0 to 240
Lower 700 MHz band 698 to 716 MHz
0 to 360
Maximum input level
Input level range
Receiver ranging
+37 dBm peak (5 W peak)
−71 to +30 dBm/1.23 MHz
Auto mode: Auto-ranges to the ideal RF power
level for the nominally expected open loop
response; provides calibrated results if actual
received power is within ±9 dB of the expected
open loop power
Manual mode: User enters expected power; if
the “active” mode is selected, the test set uses
closed loop power control to drive the mobile
to the expected power; otherwise, the mobile’s
Tx power must be within ± 9 dB of the expected
power to provide calibrated results
CDMA analyzer
Average power measurement
Input frequency ranges 411 to 484 MHz
800 to 1000 MHz
1700 to 2000 MHz
Detector types
Peak detector: In IS-95, R-RC1, and R-RC2
modes
Thermal detector: In R-RC3, R-RC4, and R-RC8
modes
Maximum input level +37 dBm peak (5 W peak)
Measurement range
−10 to +30 dBm; usable from −10 to −20 dBm
with reduced accuracy (peak detector only)
Measurement level
Auto
ranging
Measurement data
capture period
10 ms
Measurement result Average power
Concurrency support
Average power measurements can be made
concurrently with all CDMA measurements that
support concurrency
Peak detector measurement accuracy (20 to 55 °C, after calibration,
IS-95, R-RC1, or R-RC2):
–– −10 to +30 dBm:
–– 400 to 500 MHz < ± 7.3%, typically < ± 3.0%
–– 800 to 1000 MHz < ± 7.0%, typically < ± 3.0%
–– 1700 to 2000 MHz < ± 7.4%
–– RF OUT ONLY port < ± 8.3%, typically < ± 4.4%
–– −10 to −20 dBm:
–– 400 to 500 MHz typically < ± 4.4%
–– 800 to 1000 MHz typically < ± 4.2%
–– 1700 to 2000 MHz typically < ± 4.8%
–– RF OUT ONLY port is selected typically < ± 5.6%
Thermal detector measurement accuracy (accuracy with 10 internal aver-
ages; R-RC3, R-RC4, or R-RC8):
–– −10 to +30 dBm:
–– 400 to 500 MHz < ± 6.6%, typically < ± 3.0%
–– 800 to 1000 MHz < ± 6.0%, typically < ± 3.0%
–– 1700 to 2000 MHz < ± 7.2%
–– RF OUT ONLY port < ± 8.2%, typically < ± 3.3%
Measurement
repeatability
Typically < ± 0.05 dB
Measurement resolution 0.01 dBm
Zero function
Auto zeroes (no user control)