English
Language : 

33512B Datasheet, PDF (16/23 Pages) Keysight Technologies – Waveform Generators
16 | Keysight | 33500B Series Waveform Generators - Data Sheet
Specifications
Arbitrary waveform characteristics, continued
Waveform sequencing (available only on 33511B/12B/21B/22B)
Operation
Individual arbitrary waveforms (segments) can be combined into user-defined lists (sequences) to form
longer, more complex waveforms. Each sequence step specifies whether to repeat the associated segment
a certain number of times, to repeat it indefinitely, to repeat it until a trigger event occurs, or to stop and
wait for a trigger event. Additionally, the behavior of the sync output can be specified in each step.
Segment length
Sequence length
Segment repeat count
To improve throughput, up to 32 sequences totalling up to 1,024 segments can be pre-loaded into volatile
memory.
8 Sa to 1 MSa per channel (16 MSa with Option MEM) in increments of 1 sample
1 to 512 steps
1 to 1x1010, or infinite
Output characteristics
Isolation
Outputs
Signal output
Output impedance (nom)
On, off, inverted
Voltage limit
Overload protection
Amplitude
Range
Resolution
Units
Accuracy 1, 2 (spec)
DC offset
Range 3
Resolution
Units
Accuracy 1, 2 (spec)
Frequency accuracy
Connector shells for channel output(s), Sync, and Mod In are connected together but isolated from the
instrument’s chassis. Maximum allowable voltage on isolated connector shells is ±42 Vpk
50 Ω
User-selectable for each channel
User-definable VMAX and VMIN limits
Output turns off automatically when an overload is applied,
Instrument will tolerate a short-circuit to ground indefinitely
1 mVpp to 10 Vpp into 50 Ω
2 mVpp to 20 Vpp into open circuit
4 digits
Vpp, Vrms, or dBm, selectable
± 1% of setting ± 1 mVpp at 1 kHz
± (5 VDC - peak AC) into 50 Ω
± (10 VDC - peak AC) into open circuit
4 digits
VDC
± 1% of offset setting ± 0.25% of amplitude setting ± 2mV
Standard frequency reference (spec)
1 year, 23° C ± 5° C
± 1 ppm of setting ± 15 pHz
1 year, 0° C to 55° C
± 2 ppm of setting ± 15 pHz
High-stability frequency reference (spec): Option OCX
1 year, 0° C to 55° C
± 0.1 ppm of setting ± 15 pHz
1.  Add 1/10th of the output amplitude and offset accuracy specification per °C for operation at temperatures beyond 23°C ± 5°C.
2.  Auto range on.
3.  Output noise is typically 20 dB lower when (DC + Peak AC) < 320 mV (into 50 ) or 640 mV (into open circuit).