U1270 Datasheet, PDF (8/19 Pages) Agilent(Hewlett-Packard) – Agilent U1270 Series Handheld Digital Multimeters | |||
|
◁ |
Electrical specifications
DC specifications for U1271A and U1272A
Function
Voltage [1]
Resistance [2]
Current [3]
Diode Test [4]
Range
Accuracy ±(% of reading +
Resolution counts of least significant digit)
U1271A
U1272A
Test current/
Burden voltage
30 mV
0.001 mV
-
0.05 + 20
300 mV
0.01 mV
0.05 + 5
0.05 + 5
-
3V
0.0001 V
0.05 + 5
0.05 + 5
-
30 V
0.001 V
0.05 + 2
0.05 + 2
-
300 V
0.01 V
0.05 + 2
0.05 + 2
-
1000 V
0.1 V
0.05 + 2
0.05 + 2
-
Z (low input impedance)
LOW
enabled, applicable for 1000 V
-
range and resolution only.
1 + 20
30 Ω
0.001 â¦
-
0.2 + 10
0.65 mA
300 Ω
0.01 â¦
0.2 + 5
0.2 + 5
0.65 mA
3 kΩ
0.0001 kâ¦
0.2 + 5
0.2 + 5
65 mA
30 kΩ
0.001 kâ¦
0.2 + 5
0.2 + 5
6.5 mA
300 kΩ
0.01 kâ¦
0.5 + 5
0.5 + 5
0.65 mA
3 MΩ
0.0001 M⦠0.6 + 5
0.6 + 5
93 nA//10 Mâ¦
30 MΩ
0.001 Mâ¦
1.2 + 5
1.2 + 5
93 nA//10 Mâ¦
100 MΩ
-
2.0 +10
-
93 nA//10 Mâ¦
300 MΩ
0.01 Mâ¦
-
2.0 % + 10 @ <100 Mâ¦
8.0 % + 10 @ >100 Mâ¦
93 nA//10 Mâ¦
300 nS
0.01 nS
1 + 10
1 + 10
93 nA//10 Mâ¦
300 µA
0.01 µA
0.2 + 5
0.2 + 3
<0.04 V/ 100 Ω
3000 µA
0.1 µA
0.2 + 5
0.2 + 3
<0.4 V/ 100 Ω
30 mA
0.001 mA
0.2 + 5
0.2 + 3
<0.08 V/ 1 Ω
300 mA
0.01 mA
0.2 + 5
0.2 + 3
<1.00 V/ 1 Ω
3A
0.0001 A
0.3 +10
0.3 +10
<0.1 V/ 0.01 Ω
10 A
0.001 A
0.3 +10
0.3 +10
<0.3 V/ 0.01 Ω
3V
0.0001 V
0.5 + 5
0.5 + 5
Approximately 1 to 2 mA
Auto
0.0001 V
-
0.5 + 5
Approximately 0.1 to 0.3 mA
1. Notes for voltage specifications:
⢠The accuracy of the 30 to 300 mV range is specified after the Null function is used to subtract the thermal effect (by shorting the test leads).
⢠For ZLOW measurements, autoranging is disabled and the multimeterâs range is set to 1000 volts in the manual ranging mode.
2. Notes for resistance specifications:
⢠Overload protection: 1000 Vrms for short circuits with <0.3 A current.
⢠Maximum open voltage is <+3.3 V
⢠Built-in buzzer beeps when the resistance measured is less than 25 Ω ± 10 Ω. The multimeter can capture intermittent measurements longer
than 1 ms.
⢠The accuracy of the 300 Ω to 3 kΩ range is specified after the Null function is used to subtract the test lead resistance and thermal effect (by
shorting the test leads).
⢠For the ranges of 30 MΩ and 100 MΩ, the RH is specified for <60%.
⢠The accuracy for ranges <50 nS is specified after the Null function is used on an open test lead.
⢠The temperature coefficient of the 100 MΩ and 300 MΩ range is 0.1 à (specified accuracy)/°C (from -20 °C to 18 °C or 28 °C to 55 °C)
8
|
▷ |