U1270 Datasheet, PDF (10/19 Pages) Agilent(Hewlett-Packard) – Agilent U1270 Series Handheld Digital Multimeters
AC specifications for U1272A
Function
True RMS AC
Voltage [1]
Range Resolution 45 to
65 Hz
30 mV 0.001 mV
300 mV 0.01 mV
3V
0.0001 V
30 V
0.001 V
300 V
0.01 V
1000 V
0.1 V
LPF (low pass filter)
enabled, applicable for
all voltage ranges and
resolution
Z 1000 V
LOW
0.6 + 20
0.6 + 20
0.6 + 20
0.6 + 20
0.6 + 20
0.6 + 20
0.6 + 20
2.0 + 40
Accuracy
20 Hz to 1 1 kHz to
kHz
5 kHz
0.7 + 25
0.7 + 25
1.0 + 25
1.0 + 25
1.0 + 25
1.0 + 25
1.0 + 25 @
<200 Hz
5.0 + 25 @
<440 Hz
2 + 40 @
<440 Hz
1.0 + 25
1.0 + 25
1.5 + 25
1.5 + 25
1.5 + 25
1.5 + 25
-
-
5 kHz to
20 kHz
1.0 + 40
1.0 + 40
2.0 + 40
2.0 + 40
2.0 + 40
-
-
-
20 kHz
to 100
kHz
3.5 + 40
3.5 + 40
3.5 + 40
3.5 + 40
-
-
-
-
Function
True RMS
AC Current
[2]
Range
300 µA
3000 µA
30 mA
300 mA
3A
10 A
Resolution
0.01 µA
0.1 µA
0.001 mA
0.01 mA
0.0001 A
0.001 A
Accuracy
45 to
65 Hz
20 Hz
to
2 kHz
0.6 + 25 0.9 + 25
0.6 + 25 0.9 + 25
0.6 + 25 0.9 + 25
0.6 + 25 0.9 + 25
0.8 + 25 1.0 + 25
0.8 + 25 1.0 + 25
Burden
voltage/
Shunt
<0.04 V/ 100 Ω
<0.4 V/ 100 Ω
<0.08 V/ 1 Ω
<1.00 V/ 1 Ω
<0.1 V/ 0.01 Ω
<0.3 V/ 0.01 Ω
1. Notes for voltage specifications
• Overload protection: 1000 Vrms. For millivolt measurements, 1000 Vrms for short circuits with <0.3 A current.
• Input impedance: 10 MΩ (nominal) in parallel with <100 pF.
• Z impedance: 2 kΩ (nominal)
LOW
• The input signal is lower than the product of 20,000,000 V×Hz.
• For 20 to 100 kHz accuracy: Three counts of the LSD per kHz of additional error is to be added for frequencies >20 kHz and signal inputs <10%
of range.
2. Notes for current specifications
• Overload protection for 300 μA to 300 mA range: 0.44 A/1000 V; 10 × 35 mm 30 kA fast-acting fuse
• Overload protection for 3 A to 10 A range: 11 A/1000 V; 10 × 38 mm 30 kA fast-acting fuse
• Specification for 300 mA range: 440 mA continuous.
• Specification for 10 A range: 10 A continuous. Add 0.3% to the specified accuracy when measuring signals >10 to 20 A for 30 seconds
maximum. After measuring currents >10 A, cool down the multimeter for twice the duration of the measured time before proceeding with low
current measurements.
10