English
Language : 

U1270 Datasheet, PDF (14/19 Pages) Agilent(Hewlett-Packard) – Agilent U1270 Series Handheld Digital Multimeters
Frequency sensitivity for voltage measurements [1][2][3]
Input range
30 mV
300 mV
3V
30 V
300 V
1000 V
Minimum sensitivity (RMS sine wave)
15 Hz to 100 kHz 0.5 Hz to 200 kHz Up to 1 MHz
3 mV
6 mV
0.12 V
0.6 V
6V
50 V
3 mV
8 mV
0.2 V
0.8 V
8 V @ <100 kHz
50 V @ <100 kHz
-
40 mV
0.4 V
2.6 V
-
-
Trigger level for dc coupling
0.5 Hz to 200 kHz
U1271A
U1272A
-
5 mV
10 mV
15 mV
0.15 V
0.15 V
1.5 V
1.5 V
9 V @ <100 kHz 9 V @ <100 kHz
90 V @ <100 kHz 90 V @ <100 kHz
1. Maximum input for specified accuracy, refer to “AC specifications” on page 9.
2. 30 mV range applicable for U1272A only.
3. 200 kHz to 1 MHz range applicable for U1272A only.
Frequency sensitivity for current measurements [4]
Input range
300 µA
3000 µA
30 mA
300 mA
3A
10 A
Minimum sensitivity (RMS sine wave)
2 Hz to 30 kHz
100 µA
70 µA
1.2 mA
12 mA
0.12 A
1.2 A
4. Maximum input for specified accuracy, refer to “AC specifications” on page 9.
Peak hold
Signal width
Single event >1 ms
Repetitive >250 µs
Accuracy for DC Voltage and Current
Specified accuracy + 400
Specified accuracy + 1000
14