English
Language : 

U1270 Datasheet, PDF (11/19 Pages) Agilent(Hewlett-Packard) – Agilent U1270 Series Handheld Digital Multimeters
AC + DC specifications for U1272A
Function
True RMS
AC + DC
Voltage [1]
Range
30 mV
300 mV
3V
30 V
300 V
1000 V
Accuracy
Resolution 45 Hz to 20 Hz to 1 kHz to 5 kHz to
65 Hz 2 kHz 5 kHz 20 kHz
0.001 mV
0.01 mV
0.0001 V
0.001 V
0.01 V
0.1 V
0.7 + 40
0.7 + 25
0.7 + 25
0.7 + 25
0.7 + 25
0.7 + 25
0.8 + 45
0.8 + 30
1.1 + 30
1.1 + 30
1.1 + 30
1.1 + 30
1.1 + 45
1.1 + 30
1.6 + 30
1.6 + 30
1.6 + 30
1.6 + 30
1.1 + 60
1.1 + 45
2.1 + 45
2.1 + 45
2.1 + 45
-
20 kHz
to 100
kHz
3.6 + 60
3.6 + 45
3.6 + 45
3.6 + 45
-
-
Function
True RMS
AC + DC
Current [2]
Range
300 µA
3000 µA
30 mA
300 mA
3A
10 A
Accuracy
Resolution 45 Hz to 20 Hz to
65 Hz 2 kHz
0.01 µA 0.8 + 30 1.1 + 30
0.1 µA 0.8 + 30 1.1 + 30
0.001 mA 0.8 + 30 1.1 + 30
0.01 mA 0.8 + 30 1.1 + 30
0.0001 A 0.9 + 35 1.3 + 35
0.001 A 0.9 + 35 1.3 + 35
Burden
voltage/
Shunt
<0.04 V/100 Ω
<0.4 V/100 Ω
<0.08 V/1 Ω
<1.00 V/1 Ω
<0.1 V/0.01 Ω
<0.3 V/0.01 Ω
1. Notes for voltage specifications:
• Overload protection: 1000 Vrms. For millivolt measurements, 1000 Vrms for short circuits with <0.3 A current.
• Input impedance: 10 MΩ (nominal) in parallel with <100 pF.
• For 20 to 100 kHz accuracy: Three counts of the LSD per kHz of additional error is to be added for frequencies >20 kHz and signal inputs <10%
of range.
2. Notes for current specifications:
• Overload protection for 300 μA to 300 mA range: 0.44 A/1000 V; 10 × 35 mm 30 kA fast-acting fuse
• Overload protection for 3 A to 10 A range: 11 A/1000 V; 10 × 38 mm 30 kA fast-acting fuse
• Specification for 300 mA range: 440 mA continuous.
• Specification for 10 A range: 10 A continuous. Add 0.3% to the specified accuracy when measuring signals >10 to 20 A for 30 seconds
maximum. After measuring currents >10 A, cool down the multimeter for twice the duration of the measured time before proceeding with low
current measurements.
11