English
Language : 

S8980_S10810_15 Datasheet, PDF (6/10 Pages) Hamamatsu Corporation – CCD area image sensors
CCD area image sensors
S8980, S10810-11
Timing chart
Pre-integration period
AC X-ray exposure
(Trigger A)*1
DC X-ray exposure
(Trigger A)*1
Trigger B (S10810-11)*2
P1V
P2V, TG*3
P1H
P2H, SG
RG
OUT
P2V, TG
Integration period
Readout period
Tpwv
VD1
1
VD: Vertical dummy
2, 3, ... , 999, 1000, VD2
Tovr
Tpwh, Tpws
Enlarged view
P1H
P2H, SG
RG
Tpwr
OUT
S2, S3, S4, ... , S1502, S1503, S1504
D1
D2
S1
D3
D4
*1: Trigger A (S8980) is the same as AC/DC X-ray exposure form.
*2: Low active trigger pulse
*3: TG terminal can be short-circuited to P2V terminal.
KMPDC03
KMPDC0349EB
P1V, P2V, TG
P1H, P2H
SG
RG
TG-P1H
Parameter
Pulse width*15
Rise and fall times
Pulse width
Rise and fall times*15
Duty ratio
Pulse width
Rise and fall times
Duty ratio
Pulse width
Rise and fall times
Overlap time
Symbol
tpwv
tprv, tpfv
tpwh
tprh, tpfh
-
tpws
tprs, tpfs
-
tpwr
tprr, tpfr
tovr
*15: The clock pulses should be overlapped at 50% of maximum amplitude.
Min.
Typ.
Max.
Unit
30
60
-
μs
200
-
-
ns
100
500
-
ns
5
-
-
ns
-
50
-
%
100
500
-
ns
3
-
-
ns
-
50
-
%
10
50
-
ns
3
-
-
ns
18
36
-
μs
6