English
Language : 

ISE30-T1-25 Datasheet, PDF (14/21 Pages) List of Unclassifed Manufacturers – High Precision,2-color Display Digital Pressure Switch
Series ZSE30/ISE30
Specific Product Precautions 2
Be sure to read before handling.
Connection/Removal of Connector
• To connect the connector, insert it straight while
pinching the lever, and then push the lever into the jack
of the housing and lock it.
• To remove the connector, pull it straight out while
applying pressure with your thumb to the lever and
unhooking it from the jack.
Lever
DC polarity indicator
Lead wire (Brown)
Lead wire (Black)
Lead wire (Blue)
• Do not attempt to insert or pull the pressure sensor or
its connector when the power is on. A switch output
malfunction may occur.
Piping
• Cut the tube perpendicularly.
• Hold the tube and insert it into the One-touch fitting
carefully and securely all the way to the bottom.
Tube
One-touch fitting
Regulating Pressure Range and Rated Pressure Range
Caution
Set the pressure within the rated pressure range.
The regulating pressure range is the range of pressure that is possible in setting.
The rated pressure range is the range of pressure that satisfies the specifications (accuracy, linearity, etc.) on the sensor.
Although it is possible to set a value outside the rated pressure range, the specifications will not be guaranteed even if the value stays
within the regulating pressure range.
Switch
–100 kPa
0
Pressure range
100 kPa
500 kPa
1 MPa
For vacuum/
low pressure
–100 kPa
ZSE30
–101 kPa
100 kPa
101 kPa
For positive
pressure
ISE30 –100 kPa
0
(–0.1 MPa)
1 MPa
1 MPa
Rated pressure range of switch
Regulating pressure range of switch
16-2-14