English
Language : 

PSPRL1103M Datasheet, PDF (3/7 Pages) Wall Industries,Inc. – Medical AC/DC Switching Power Supplies
Rev. A
PSPRL1103M Series
150~300 Watts
PFC, Single Outputs
Medical AC/DC Switching Power Supplies
Model Number (1)
PSPRL1103MU-12
PSPRL1103MU-15
PSPRL1103MU-24
PSPRL1103MU-36
PSPRL1103MU-48
Model Number (1)
PSPRL1103ME-12
PSPRL1103ME-15
PSPRL1103ME-24
PSPRL1103ME-36
PSPRL1103ME-48
Model Number (1)
PSPRL1103MF-12
PSPRL1103MF-15
PSPRL1103MF-24
PSPRL1103MF-36
PSPRL1103MF-48
Input Voltage Range
90 ~ 264 VAC
MODEL SELECTION TABLES
U-CHASSIS MODELS (TYPE “U”)
Output Voltage (1)
12 ~ 13.8 VDC
Preset Voltage (1)
12 VDC
Output Current
Convection Forced Air
12.5A
25A
14 ~ 16 VDC
15 VDC
10A
20A
23 ~ 28 VDC
24 VDC
6.25A
12.5A
35 ~ 43 VDC
36 VDC
4.17A
8.33A
44 ~ 52 VDC
48 VDC
3.125A
6.25A
ENCLOSED WITH REAR-SIDE BUILT-IN FAN MODELS (TYPE “E”)
Input Voltage Range
Output Voltage (1)
Preset Voltage (1)
Output Current
12 ~ 13.8 VDC
12 VDC
25A
14 ~ 16 VDC
15 VDC
20A
90 ~ 264 VAC
23 ~ 28 VDC
24 VDC
12.5A
35 ~ 43 VDC
36 VDC
8.33A
44 ~ 52 VDC
48 VDC
6.25A
ENCLOSED WITH TOP-SIDE BUILT-IN FAN MODELS (TYPE “F”)
Input Voltage Range
Output Voltage (1)
Preset Voltage (1)
Output Current
12 ~ 13.8 VDC
12 VDC
25A
14 ~ 16 VDC
15 VDC
20A
90 ~ 264 VAC
23 ~ 28 VDC
24 VDC
12.5A
35 ~ 43 VDC
36 VDC
8.33A
44 ~ 52 VDC
48 VDC
6.25A
Output Power (2)
Convection Forced Air
150W
300W
150W
300W
150W
300W
150W
300W
150W
300W
Output Power (2)
300W
300W
300W
300W
300W
Output Power (2)
300W
300W
300W
300W
300W
NOTES
1. All output ranges are covered in agency certifications and the preset voltage will be set as standard models if nothing different is requested. If desired
preset output does not appear, please contact factory.
2. PSPRL1103MU Models (U-Chassis): Needs 25CFM min. forced airflow to achieve 300W maximum power.
PSPRL1103ME Models (Enclosed with rear-side built-in fan): 300W max. with built-in fan airflow.
PSPRL1103MF Models (Enclosed with top-side built-in fan): 300W max. with built-in fan airflow.
3. Provides peak power to 600W within 500µs for all models; for longer duty duration must contact manufacturer.
4. 1% minimum load is required to maintain the ripple and regulation specifications.
5. Output is fully isolated.
DERATING CURVE
120
100
80
60
40
20
0
0 10 20 30 40 50 60 70 80
Ambient Temperature (°C)
06/22/2012
Wall Industries, Inc. 5 Watson Brook Road Exeter, NH 03833
603-778-2300 www.wallindustries.com Fax 603-778-9797
Page 3 of 7