English
Language : 

PSMAD100 Datasheet, PDF (2/8 Pages) Wall Industries,Inc. – AC/DC Power Supplies
Wall Industries, Inc.
Rev C
PSMAD100 SERIES
100 Watt, Medical AC/DC Power Supplies
SPECIFICATIONS: PSMAD100 SERIES
All specifications are based on 25°C, Nominal Input Voltage, and Maximum Output Current unless otherwise noted.
We reserve the right to change specifications based on technological advances.
SPECIFICATION
TEST CONDITIONS
Min Typ Max Unit
INPUT SPECIFICATIONS
Operating Input Voltage Range
AC input
DC input
85
264
VAC
120
370
VDC
Input Frequency
AC input
47
63
Hz
Input Current
115VAC and full load
230VAC and full load
1.15
0.55
A
No load Input Power
230VAC
0.3
W
Input Inrush Current
230VAC
60
A
Power Factor Correction
0.95
Input Protection
Internal fuse in line and neutral
T3.15A / 250VAC
OUTPUT SPECIFICATIONS
Output Voltage
See Table
Initial Set Voltage Accuracy
230VAC and full load
-1.0
+1.0
%
Line Regulation
Low line to high line at full load
-0.2
+0.2
%
Load Regulation
No load to full load
10% load to 90% load
-0.5
-0.4
+0.5
+0.4
%
Voltage Adjustability
-10
+10
%
Output Power
100
W
Output Current
See Table
Minimum Load
0
%
Ripple & Noise (20MHz BW)
Transient Response
With 10µF/25V 1206 X7R MLCC capacitor
With 10µF/25V 1206 X7R MLCC capacitor
With 1µF/50V 1206 X7R MLCC capacitor
With 1µF/50V 1206 X7R MLCC capacitor
With 1µF/50V 1206 X7R MLCC capacitor
With 0.1µF/100V 1206 X7R MLCC capacitor
Load step from 50~75% change at 2.5A/µs
12V output model
15V output model
24V output model
28V output model
36V output model
48V output model
Peak Deviation
Recovery Time
120
150
160
mVp-p
180
190
340
3
% Vout
500
µs
Start-Up Time
1000
ms
Rise Time
20
ms
Hold-up Time
115VAC and full load
16
ms
Temperature Coefficient
-0.02
+0.02 %/°C
PROTECTION
Over Voltage Protection
% of Vout (nom); latch mode
115
130
%
Over Load Protection
% of Iout rated; hiccup mode
115
150
%
Short Circuit Protection
Continuous, automatic recovery
GENERAL SPECIFICATIONS
Switching Frequency
60
kHz
Input to Output
4000
Isolation Voltage
1 minute (2MOPP insulation)
Input to FG
1500
VAC
Output to FG
1500
Isolation Resistance
500VDC
0.1
GΩ
Leakage Current
264VAC
75
µA
05/22/2013
Wall Industries, Inc. • 5 Watson Brook Road, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 • Fax 603-778-9797
website: www.wallindustries.com • e-mail: sales@wallindustries.com