English
Language : 

OA1MPA Datasheet, PDF (19/28 Pages) STMicroelectronics – Medical instrumentation
OA1MPA, OA2MPA, OA4MPA
Application information
4.6
Initialization time
The OA1MPA, OA2MPA, and OA4MPA series of devices use a proprietary trimming
topology that is initiated at each device power-up and allows excellent Vio performance to be
achieved. The initialization time is defined as the delay after power-up which guarantees
operation within specified performances. During this period, the current consumption (ICC)
and the input offset voltage (Vio) can be different to the typical ones.
Figure 22. Initialization phase
The initialization time is VCC and temperature dependent. Table 7 sums up the
measurement results for different supply voltages and for temperatures varying from -40 °C
to 125 °C.
VCC (V)
1.8
3.3
5
Table 7. Initialization time measurement results
Temperature: -40 °C
Temperature: 25 °C
Temperature: 125 °C
Tinit (ms)
37
2.9
2.4
ICC phase 1 (mA)
0.33
1.4
3.2
Tinit (ms)
3.2
0.95
0.85
ICC phase 1 (mA)
0.40
1.3
2.4
Tinit (ms)
0.35
0.34
0.31
ICC phase 1 (mA)
0.46
1.2
2.9
4.7
PCB layouts
For correct operation, it is advised to add a 10 nF decoupling capacitors as close as
possible to the power supply pins.
DocID025992 Rev 1
19/28
28