English
Language : 

MB60S Datasheet, PDF (4/9 Pages) SL Power Electronics – 60W Single Output Medical Series
MB60S
60W Single Output
Medical Series
EMI/EMC Compliance
PARAMETER
Conducted Emissions:
SPECIFICATION
EN55011/22 Class B; FCC Part 15
NOTES
Radiated Emissions:
Harmonic Current
Emissions
Voltage Fluctuations &
Flicker
Static Discharge
Immunity:
RF Field Susceptability
Fast Transients/Bursts
Surge Susceptability
Conducted RF
Susceptability
Power Frequency
Magnetic Field Test
Voltage Sags & Surges
EN55011/22 Class A; FCC Part 15
EN61000-3-2, Class A
EN61000-3-3
EN61000-4-2 6kV contact, 8kV air, Criteria A
EN61000-4-3 (3V/m), Criteria A
EN61000-4-4 (PS: 2kV-40A, other lines 1kV-
20A), Criteria B
EN61000-4-5, Installation Class 3 (1kV diff.
mode, 2kV common mode), Criteria A
EN61000-4-6 (3Vrms), Criteria A
EN61000-4-8 (3A/m), Criteria A
EN61000-4-11,
95% dip/0.5 cycle (Criteria A),
60%/5cycles (Criteria B ), 30%/25 cycles
(Criteria A).
Performance criteria are defined as
following:
A – Normal performance
during and after the test
B – Temporary degradation,
self-recoverable
C – Temporary degradation,
operator intervention required
to recover the operation
Notes:
1. Specifications subject to change without notice.
2. Specifications are for convection rating at factory settings with 115Vac input and 25°C ambient unless
otherwise stated.
www.slpower.com
MB60S 6-July-15
4