English
Language : 

MB60S Datasheet, PDF (2/9 Pages) SL Power Electronics – 60W Single Output Medical Series
MB60S
60W Single Output
Medical Series
Leakage Current
(Input–Earth):
Leakage Current
(Output–Earth):
Leakage Current
(Input-Output):
Input Fuses:
Efficiency
MB60S12K
MB60S15K
MB60S18K
MB60S24K
MB60S36K
MB60S48K
No Load Input Power:
Turn-on Time:
Hold-up Time:
<275µA@264Vac, 60 Hz input, NC
N/A
<90µA@264Vac, 60 Hz input, NC
F1, F2: 4A, 250VAC
Typical
83%
85%
85%
88%
88%
88%
<0.5W
<2 Seconds at 120Vac.
16mS minimum from loss of ac input at
120 Vac, full load.
IEC 60601-1 3rd Ed – 8.7.3.c
Fuses provided on all models
Measured at 120Vac and full load
24V, 36V, and 48V Models meet Level V
requirement
Meet Level V, standby Power Consumption
55 Watts for 12V output
DC Output Specifications
PARAMETER
SPECIFICATION
NOTES
Output Power:
60W continuous for operation from -10ºC
to 50ºC 55 Watts for 12V output.
Cooling:
Total Regulation:
Overload Protection:
Short Circuit
Protection:
Overvoltage
Protection:
Overtemperature
Protection:
Minimum Load:
Ripple and Noise:
Transient Response:
Overshoot:
Convection
±2% for all models
120% - 180% of rated output current
value, Hiccup Mode
Short across the output terminals will not
cause damage to the unit. Hiccup Mode
OVP firing reduces output voltage to
<50% of nominal in <50mS. See chart for
trip range
Automatic Power Shutdown at Tc =
155ºC,
No minimum load is required
0.5% RMS, 1% pk-pk for all models.
500µs typ. response time for return to
within 0.5% of final value for a 50% load
change, ∆i/∆t< 0.2A/µs. Max. voltage
deviation is 3.5%.
5% overshoot at turn-on, 5% overshoot at
turn-off, under all conditions.
Total regulation is the maximum deviation from
nominal voltage for all loading conditions
20 MHz Bandwidth, differential mode.
Measured with noise probe directly across
output terminals, and load terminated with
0.1µF ceramic and 10µF low ESR capacitors
www.slpower.com
MB60S 6-July-15
2