MB60S Datasheet, PDF (3/9 Pages) SL Power Electronics – 60W Single Output Medical Series
MB60S
60W Single Output
Medical Series
Safety Standard Compliance
Agency
UL
CSA
Demko
CB Report
Isolation Type:
Isolation Specifications
PARAMETER
CONDITIONS
Insulation Safety
Rating:
Input to Ground
Input to Output
Output to Ground
Electric Strength Test
Voltage:
Input to Ground
Input to Output
Output to Ground
CONDITIONS
ANSI/AAMI ES60101:2005, 3rd Edition
CAN/CSA-C22.2 No. 60601-1 (2008)
EN 60601-1:2006
IEC 60601-1 (3rd Edition)
B rated
Rating
1 MOPP
2 MOPP
Functional
1800Vac
4000Vac
500Vac
NOTES
.
Environmental Specifications
PARAMETER
SPECIFICATION
Operating
Temperature:
Temperature Derating:
Cooling:
Storage Temperature:
Altitude:
Relative Humidity:
Shock:
Vibration:
-10°C to +80°C
For Output Voltage ≥24V, derate output power to
50W @ 60C, 40 Watt @ 70C, and 20 Watts for 80C
Convection
-40°C to +85°C
Operating: -500 to 3,000 meter
Non-operating: -500 to 40,000 ft.
5% to 95%, non-condensing
Non-Operating: Half-sine, 40 gpk, 10mS, 3 axes, 6
shocks total
Random vibration per MIL-STD-810E, Method
514.4, Cat. 1, Figure 514.4-1, 1 hr in each of three
axes
NOTES
-40°C Startup guaranteed
<24V will derate to 40W at 60C,
30W at 70C, and 20W at 80C
Reliability Specifications
PARAMETER
SPECIFICATION
MTBF:
700,000 hours, 25°C ambient, full load
Warranty:
HALT Data:
3 Years
Per SL Power Halt procedure
NOTES
Calculation is done based on
Telcordia. Reports for each
model is available
Limited
Report is available
www.slpower.com
MB60S 6-July-15
3