FMO Series
Mass flow sensors for gases
GAS CORRECTION FACTORS7
Gas type
Helium (He)
Hydrogen (H2)
Argon (Ar)
Nitrogen (N2)
Oxygen (O )
2
Air
Nitric oxide (NO)
Carbon monoxide (CO)
Methane (CH4)
Ammonia (NH3)
Nitrous oxide (N O)
2
Nitrogen dioxide (NO2)
Carbon dioxide (CO2)
Correction factor (approx.)
0.58
0.78,9
0.95
1.0
1.0
1.0
1.0
1.0
1.1
1.1
1.35
1.35
1.35
Notes:
7 Gas correction factors are referenced to nitrogen (N2) as calibration gas type. Approximate gas correction factors are provided
as guidelines only. Individual gas types may perform differently at temperature extremes and varying flow rates.
8 When sensing Hydrogen (H2) or Helium (He) it may be necessary to power the mass flow sensor using increased supply voltage:
Hydrogen typ. 12 V, Helium typ. 15 V
9 Hydrogen (H2) flow measurement requires the use of a special sensor. These devices provide normal operation when sensing
hydrogen flow and are designated with an "H" at the end of the order number.
ORDERING INFORMATION - AVAILABLE LISTINGS
Flow range
±25 sccm
±1000 sccm
Dry gas
-
FMOL001DB
Hydrogen gas9
FMOM025HB
-
Sensortechnics PRO services:
· Extended guarantee period of 2 years
· Improved performance characteristics
· Custom product modifications and adaptations even for small quantities
· Advanced logistics models for supply inventory and short delivery times
· Technical support through application engineers on the phone or at your site
· Fastest possible technical response for design and QA engineers
... plus other services on request
Sensortechnics reserves the right to make changes to any products herein. Sensortechnics does not assume any liability arising out of the
application or use of any product or circuit described herein, neither does it convey any license under its patent rights nor the rights of others.
E / 11723 / A
6/6
www.sensortechnics.com