English
Language : 

MAX11359A Datasheet, PDF (36/66 Pages) Maxim Integrated Products – 16-Bit Data-Acquisition System with ADC, DAC, UPIOs, RTC, Voltage Monitors, and Temp Sensor
16-Bit Data-Acquisition System with ADC, DAC,
UPIOs, RTC, Voltage Monitors, and Temp Sensor
Table 6. Setting the ADC Conversion Rate*
CONTINUOUS
CONVERSION
RATE (sps)
10
40
50
60
200
240
400
512
SINGLE
CONVERSION
RATE (sps)
2.5
10
12.5
15
50
60
100
128
RATE2
0
0
0
0
1
1
1
1
RATE1
0
0
1
1
0
0
1
1
RATE0
0
1
0
1
0
1
0
1
The actual rates are:
NOMINAL
CONTINUOUS
CONVERSION
RATE (sps)
10
40
50
60
200
240
400
512
DECIMATION
RATIO
1096
274
220
183
55
46
27
23
ACTUAL
CONTINUOUS
CONVERSION
RATE (sps)
10.01042142
40.04168568
49.87009943
59.953125
199.4803977
238.5091712
406.3489583
477.0183424
*Calculate the ADC sampling rate using the following
equation:
fS
=
448
×
fHFCLK
decimation
ratio
where fHFCLK = 4.9152MHz nominally.
RATE<2:0>: ADC conversion-rate-setting bits. These
three bits set the conversion rate of the ADC as shown
in Table 6. The initial conversion requires four conver-
sion cycles for valid data, and subsequent conversions
require only one cycle (if CONT = 1). A full-scale input
change can require up to five cycles for valid data if
the digital filter is not reset with the STRT or S bit.
MODE<2:0>: Conversion-mode bits. These three bits
determine the type of conversion for the ADC as shown
in Table 7. When the ADC finishes an offset calibration
and/or gain calibration, the MODE<2:0> bits clear to 0
hex, the ADD bit in the STATUS register asserts, and
an interrupt asserts on INT (or UPIO_ if programmed as
DRDY) if MADD is unmasked. Perform a gain calibra-
tion after achieving the desired offset (calibrated or
not). If an offset and gain calibration are performed
together (MODE<2:0> = 7 hex), the offset calibration is
performed first followed by the gain calibration, and the
µC is interrupted by INT (or UPIO_ if programmed as
DRDY) if MADD is unmasked only upon completion of
both offset and gain calibration. After power-on or cali-
bration, the ADC does not begin conversions until initi-
ated by the user (see the ADCE and STRT bit
descriptions in this section and see the S bit descrip-
tions in the MUX Register section). See the GAIN CAL
Register and OFFSET CAL Register sections for details
on system calibration.
Table 7. Setting the ADC Conversion Mode
CONVERSION MODE
Normal
System Offset Calibration
System Gain Calibration
Normal
Normal
Self Offset Calibration
Self Gain Calibration
Self Offset and Gain
Calibration
MODE2
0
0
0
0
1
1
1
MODE1
0
0
1
1
0
0
1
MODE0
0
1
0
1
0
1
0
1
1
1
36 ______________________________________________________________________________________