English
Language : 

U1401B Datasheet, PDF (8/10 Pages) Keysight Technologies – Multi-function Calibrator/Meter
08 | Keysight | U1401B Handheld Multi-function Calibrator/Meter - Data Sheet
General Speciications
Display
Power supply
Power consumption
Battery life
Charging time
Measurement rate
Common Mode Rejection Ratio (CMRR)
Normal Mode Rejection Ratio (NMRR)
Operating environment
Storage environment
Altitude
Safety compliance
EMC compliance
Temperature coeficient
Dimensions (H x W x D)
Weight
Calibration
Warranty
Both primary and secondary displays are 5-digit on the liquid crystal display (LCD) with a maximum
resolution of 50,000 counts and automatic polarity indication. Backlight available.
– 9.6 V Ni-MH rechargeable batteries: 1.2 V x 8 pieces. No cadmium, lead or mercury.
– External switching adapter: AC 100 V to 240 V, 50/60 Hz input and DC 24 V/2.5 A output.
– Battery charging: 9.3 VA typical
– Sourcing of constant current at 25 mA, maximum load: 5.5 VA typical on 24 V DC adapter, 2.4 VA
typical on 9.6 V batteries
– Meter only: 1.8 VA typical on 24 V DC adapter, 0.6 VA typical on 9.6 V batteries
Assuming fully-charged Ni-MH batteries:
Meter only: 20 hours (approximately)
Source/Meter: 4 hours (approximately)
will appear when voltage drops below 9 V (approximately)
Three hours (approximately) in 10 °C to 30 °C environment
NOTE: Prolonged charging is required if battery is fully discharged
Three readings/second, except for:
– AC+DC: 1 reading/second
– Frequency and duty cycle (> 1 Hz): 1 reading/second
– Pulse width (> 1 Hz): 0.25 to 1 reading/second
> 90 dB at DC, 50/60 Hz ± 0.1% (1 kΩ unbalanced)
> 60 dB at DC, 50/60 Hz ± 0.1%
0 °C to 40 °C; up to 80% relative humidity (R.H.) for temperatures up to 31 °C, decreasing linearly to
50% R.H. at 40 °C
–20 °C to 60 °C with batteries removed; 5% to 80% R.H. non-condensing
0 to 2000 m
IEC 61010-1:2001/EN61010-1:2001 (2nd Edition), CAN/CSA-C22.2 No. 61010-1-04,
ANSI/UL 61010-1:2004, CAT II 150 V Overvoltage Protection, Pollution Degree 2
IEC61326-2-1:2005/EN61326-2-1:2006, ICES-001:2004, AS/NZS CISPR11:2004
– Input: 0.15 x (speciied accuracy)/°C (from 0 °C to 18 °C or 28 °C to 40 °C)
– Output: ± (50 ppm output + 0.5 digit)/°C
192 mm x 90 mm x 54 mm
0.98 kg with holster and batteries
One-year calibration cycle recommended
– Three years for main unit
– Three months for standard accessories unless otherwise speciied