U1401B Datasheet, PDF (5/10 Pages) Keysight Technologies – Multi-function Calibrator/Meter
05 | Keysight | U1401B Handheld Multi-function Calibrator/Meter - Data Sheet
Input Speciications
Temperature speciications
Thermocouple type
K
Range
–40 °C to 1372 °C
–40 °F to 2502 °F
Resolution
0.1 °C
0.1 ºF
Accuracy1
0.3% + 3 °C
0.3% + 6 °F
Overload protection
250 Vrms
1. Accuracy is speciied for meter operation only, excludes thermocouple probe tolerance and with the instrument placed in the operating area for at least one
hour.
Resistance speciications
Range
500 W
5 kW
50 kW
500 kW
5 MW
50 MW
Resolution
0.01 W
0.1 W
1W
10 W
0.1 kW
1 kW
Accuracy
0.15% + 82
0.15% + 52
1% + 83
Minimum input current
0.45 mA
0.45 mA
45 µA
4.5 µA
450 nA
45 nA
2. Accuracy is speciied after applying the Relative function to offset any test lead resistance and thermal effect.
3. Accuracy is speciied for < 60% R.H.
Overload protection
250 Vrms
Diode and continuity speciications
For diode test, the overload protection is 250 Vrms and the instrument will beep when the reading is below 50 mV (approximately). For
continuity test, the instrument will beep when the resistance is less than 10.00 Ω.
Resolution
0.1 mV
Accuracy
0.05% + 5
Test current
Approximately 0.45 mA
Open voltage
< +4.8 VDC
1 ms peak hold speciications
Signal width
Single event > 1 ms
Accuracy for DC mV/voltage/current
2% + 400 for all ranges