07 | Keysight | U1401B Handheld Multi-function Calibrator/Meter - Data Sheet
Output Speciications
Accuracy is given as ± (% of output + counts of least signiicant digit (LSD)) at 23 °C ± 5 °C, with relative humidity less than 80% R.H.
and after a warm-up period of at least ive minutes. The maximum input voltage protection is 30 VDC.
Constant voltage and current outputs
Function
Constant voltage (CV)
Constant current (CC)
Range
± 1.5000 V
± 15.000 V
± 25.000 mA
Resolution
0.1 mV
1 mV
1 µA
Accuracy
0.03% + 3
0.03% + 5
Maximum output
25 mA or above1
12 V or above2, 3
1. Loading coeficient: 0.012 mV/mA for 1.5 V output.
2. Loading coeficient: 1 µA/ V. The minimum output voltage is based on 20 mA into a 600 Ω load.
3. If the current loop has a 24-V power, a minimum output voltage of 24 V is achievable with a 20 mA current into a 1200-Ω load (applicable only in Simulation
Mode).
Square wave output
Output
Frequency (Hz)
Duty Cycle (%)4
Pulse Width (ms)4
Amplitude (V)
Range
0.5, 1, 2, 5, 10, 15, 20, 25, 30, 40, 50,
60, 75, 80, 100, 120, 150, 200, 240,
300, 400, 480, 600, 800, 1200, 1600,
2400, 4800
0.39% to 99.60%
1/Frequency
5 V, 12 V
±5 V, ±12 V
Resolution
0.01
0.390625%
Range/256
0.1 V
0.1 V
Accuracy
0.005% + 1
0.01% + 0.2%5
0.01% + 0.3 ms
2% + 0.2 V
2% + 0.4 V
4. The positive or negative pulse width must be greater than 50 µs to enable adjustment of duty cycle or pulse width under different frequencies. Otherwise,
the accuracy and range will be different from the speciications deined.
5. For signal frequencies greater than 1 kHz, add an addition of 0.1% per kHz.