Performance characteristics1 (continued)
Vertical system: oscilloscope channels (continued)
Single-cursor accuracy
±{DC vertical gain accuracy + DC vertical offset accuracy + 0.2% full scale (~1/2 LSB)}
Example:
For 50 mV signal, scope set to 10 mV/div (80 mV full scale), 5 mV offset,
Accuracy = ±{4.0% (80 mV) + 0.1(10 mV) + 2.0 mV + 0.5% (5 mV) + 0.2% (80 mV)}
= ±6.385 mV
Dual-cursor accuracy
±{DC vertical gain accuracy + 0.4% full scale (~1 LSB)}
Example:
For 50 mV signal, scope set to 10 mV/div (80 mV full scale), 5 mV offset,
Accuracy = ±{4.0% (80 mV) + 0.4% (80 mV)}
= ±3.52 mV
Horizontal
Range
1 ns/div to 50 s/div
Time base accuracy
20 ppm
Delay range
Pre-trigger: –100 %
Post-trigger: +100 %
Modes
Main, roll, XY
XY
Yes
Reference position
Center
Trigger System
Sources
Ch 1, Ch 2, Ext (not applicable for TV trigger)
Modes
Normal, single, auto trigger
Hold off time
60 ns
Selections
Edge, pulse width, TV
–– Edge
Triggers on a rising or falling edge, alternating, or either edge of any source
–– Pulse width
Triggers on a pulse width greater than, equal to, or less than a specified time limit, with time limits ranging
from 16 ns to 10 s
Minimum lower limit: 8 ns
Minimum upper limit: 16 ns
Maximum pulse width setting: 10 s
–– TV
Triggers on one of three standard television waveforms: NTSC, PAL, SECAM
TV trigger sensitivity: 0.6 division of sync signal. Modes supported include Field 1, Field 2, all fields, or any
line within a field
Autoscale
Single-button automatic setup of all channels
Oscilloscope channel triggering
Range (internal)
±4 div from center screen
Coupling
AC (< 15 Hz)
LF reject (~ 35 kHz)
HF reject (~ 35 kHz)
External (EXT) triggering
Input impedance
1 MW: ≈ 16 pF
Maximum input
CAT I 30 Vrms, 42 Vpk
Range
DC coupling: trigger level ±1.25 V and ±2.5 V
EXT trigger pulse width
> 2.5 ns
1.  All characteristics are typical performance values and are not warranted. Characteristics are valid after a 30-minute warm-up period and within ±10 ºC of
last calibration temperature.