U1250 Datasheet, PDF (6/15 Pages) Agilent(Hewlett-Packard) – Agilent U1250 Series Handheld Digital Multimeters
6 | Keysight | U1250 Series Handheld Digital Multimeter - Data Sheet
U1252B/U1253B AC current specifications
Function
True RMS AC
current 7
Range
Resolution
500.00 µA 3
5000.0 µA
50.000 mA
440.00 mA
5.0000 A
10.000 A 4
0.01 µA
0.1 µA
0.001 mA
0.01 mA
0.0001 A
0.001 A
Accuracy ± (% of reading + no. of least significant digit)
Frequency
20 Hz ~ 45 Hz
45 Hz ~ 1 Khz
1 Khz ~ 20 Khz 6
1.0+20
0.7+20
0.75+20
1.5+20 5
1.5+20
3+60, <3 A/5 kHz
20 Khz~100 Khz 6 8
5.0+80
N/A
U1252B/U1253B AC+DC voltage specifications
Function
True RMS AC
voltage 1 2
Range
Resolution
50.000 Mv
500.00 mV
1000.0 mV
5.0000 V
50.000 V
500.00 V
1000.00 V
0.001 Mv
0.01 mV
0.1 mV
0.0001 V
0.001 V
0.01 V
0.1 V
Accuracy ± (% of reading + no. of least significant digit)
Frequency
30 Hz ~ 45 Hz 45 Hz ~ 1 kHz 1 kHz ~ 5 kHz 5 kHz ~ 15 kHz
1.5+80
0.4+60
0.7+60
0.8+60
1.5+65
0.4+30
0.4+30
0.8+45
15 kHz~100 kHz 8
3.5+220
3.5+125
0.6+30
1.5+45
0.4+30
N/A
N/A
0.4+45
0.4+45
N/A
N/A
U1252B/U1253B AC+DC current specifications
Function
True RMS AC
current
Range
Resolution
500.00 µA 3
5000.0 µA
50.000 mA
440.00 mA
5.0000 A
10.000 A 4
0.01 µA
0.1 µA
0.001 mA
0.01 mA
0.0001 A
0.001 A
Accuracy ± (% of reading + no. of least significant digit)
Frequency
30 Hz ~ 45 Hz
45 Hz ~ 1 kHz 6
1 kHz ~ 20 kHz 6
1.1+25
0.8+25
0.8+25
1.2+25
0.9+25
0.9+25
1.8+30 5
0.9+30
0.9+25
3.3+70, <3 A/5 kHz
1.  Input impedance: Refer to Table A on page 10.
2.  ACV measurement is AC coupled; while ACmV and AC μA/mA/A measurement are DC coupled. All
specifications are valid from 5% to 100% of range. The crest factor may be up to 3 at full scale, up to
5 at half-scale, except for 1000 mV and 1000 V ranges, where the crest factor is 1.5 at full scale and 3
at half scale.
3.  Input current >35 µARMS.
4.  Current can be measured from 2.5 A up to 10 A continuously. An additional 0.5% needs to be added to
the specified accuracy if the signal measured is in the range of 10 A ~ 20 A for 30 seconds maximum.
After measuring a current of >10 A, leave the meter to cool down for twice the measuring time used
before application of low current measurement.
5.  Input current < 3 ARMS.
6.  These specifications are for typical performance.
7.  Crest factor ≤ 3.0 at full scale, 5.0 at half scale except the 1000 mV and 1000 V ranges where it is
1.5 at full scale, 3.0 at half scale. For non-sinusoidal waveform, add 0.1 % of reading ± 0.3 % of range.
8.  Additional error to be added for frequency > 15 kHz and signal input < 10 % of range: 3 counts of LSD per kHz.
9.  2% over-range on all ranges except AC 1000 V range