81100 Datasheet, PDF (6/13 Pages) Keysight Technologies – Pulse Pattern Generators
Level/Pulse Performance Characteristics
Level specifications are valid after a 5 ns (Agilent 81112A) or 30 ns (Agilent 81111A/Agilent 81105A) typical settling time.
Mainframe
Amplitude
Level window
Accuracy
Resolution
Output connectors
Source impedance
Accuracy
Max. external voltage
Short circuit current
50 Ω into 50 Ω
50 Ω into 50 Ω
1 kΩ into 50 Ω
50 Ω into 50 Ω
1 kΩ into 50 Ω
50 Ω into 50 Ω
1 kΩ into 50 Ω
Dynamic crosstalk
Base line noise
Overshoot/preshoot/ringing
Agilent 81101A
Agilent 81105A
100 mVpp to 10.0 Vpp
-10.0 V to +10.0 V
-20.0 V to +20.0 V
± (3% + 75 mV)
± (3% + 150 mV1
10 mV
20 mV
BNC single-ended
Selectable 50 Ω or 1 kΩ
Typ. ± 1%
± 24 V
±400 mA max.
(doubles for channel addition)
< 0.1% typ.
10 mV RMS typ. 4 mV RMS typ.
± 5% of amplitude ± 20 mV
Agilent 81110A
Agilent 81111A
Agilent 81110A
Agilent 81112A
100 mVpp to 10.0 Vpp
-10.0 V to +10.0 V
-20.0 V to +20.0 V
100 mVpp to 3.8 Vpp
- 2.0 V to 3.8 V
N/A
± (1% + 50 mV)
± (1% + 100 mV)1
± (2% + 50 mV)
N/A
10 mV
20 mV
10 mV
N/A
BNC single-ended
BNC differential
Selectable 50 Ω or 1 kΩ
50 Ω only
Typ. ± 1%
Typ. ± 1%
± 24 V
-2.2 V to +5.5 V
±400 mA max.
-84 mA to + 152 mA
(doubles for channel addition)
< 0.1% typ.
< 0.1% typ.
10 mV RMS typ. 4 mV RMS typ. 10 mV RMS typ. 4 mV RMS typ.
± 5% of amplitude ± 20 mV ± 5% of amplitude ± 50 mV
1. In ± 19 V level window.
Level parameters: voltage or current, high
or low level, offset or amplitude.
On/off: relays connect/ disconnect output
(HiZ).
Load compensation: the actual load value
can be entered (forloads ≠ 50 Ω) to display
actual output values. (Applies to the
Agilent 81105A and Agilent 81111A only).
Normal/complement: selectable.
Limit: programmable high and low levels
can be limited to protect the device-under-
test.
Channel Addition (with Agilent 81105A or Agilent 81111A output channels)
If the instrument is equipped with 2 output modules, channel 2 can be added to channel 1 internally. In this case the second output is
disabled. The additional fixed delay on the second channel is typ. 2.5 ns. The following parameters differ from the above specifications if
two output modules (Agilent 81105A/Agilent 81111A) are added.
Mainframe
Amplitude
Source impedance
Level window
Max. frequency
Min. transitions
50 Ω into 50 Ω
1 kΩ into 50 Ω
50 Ω into 50 Ω
1 kΩ into 50 Ω
50 Ω into 50 Ω
1 kΩ into 50 Ω
50 Ω into 50 Ω
1 kΩ into 50 Ω
Agilent 81104A with two
Agilent 81105A output modules
100 mVpp tp 20.0 Vpp
200 mVpp to 20.0 Vpp
Selectable from 50 Ω or 1 kΩ
-20.0 V +20.0 V
-20.0 V to +20.0 V
60 MHz typ.
15 MHz typ.
2 ns typ. (channel one) 5 ns typ. (channel two)
20 ns typ. both channels
Agilent 81110A with two
Agilent 81111A output modules
100 mVpp tp 20.0 Vpp
200 mVpp to 20.0 Vpp
Selectable from 50 Ω or 1 kΩ
-20.0 V +20.0 V
-20.0 V to +20.0 V
60 MHz typ.
15 MHz typ.
2 ns typ. (channel one) 5 ns typ. (channel two)
20 ns typ. both channels
6
Data Sheet 81100 Family of Pulse Pattern Generators