U2352A Datasheet, PDF (21/25 Pages) Keysight Technologies – USB Modular Multifunction Data Acquisition Devices | |||
|
◁ |
System accuracy specifications
The U2802A system accuracy specifications are shown in Table 1, Table 2, and Table 3. These measurements are derived from
the U2802A and DAQ input accuracy specifications without including the thermocouple error. Refer to âCalculating System
Accuracyâ section in the Keysight U2802A 31-Channel Thermocouple Input Device Userâs Guide for calculation methodology.
Table 1. Measurement accuracy of the U2355A and U2356A at 23 °C ± 5 °C
T/C type
B
E
J
K
N
R
S
T
Thermocouple measurement accuracy (U2355A, U2356A at 23 °C ± 5 °C)
ITS-90 Temperature range (°C)
Low
High
0
1820
â270
1000
â210
1200
â270
1372
â270
1300
â50
1768
â50
1768
â270
400
Optimum measurement
range(°C)
Low
High
1100
1820
400
1100
â150
1000
â200
â150
â150
1200
â210
â150
â100
1200
â200
â100
â100
1300
â200
â100
300
1760
â50
300
400
1760
â50
400
â100
400
â200
â100
Without
averaging
(± °C)
1.9
4.4
1.7
2.4
1.6
2.7
1.5
2.7
1.5
3.0
2.0
5.0
2.1
4.5
1.5
2.7
50 points
averaging
(± °C)
1.2
2.5
1.6
2.3
1.5
2.6
1.4
2.6
1.3
2.7
1.4
3.1
1.6
2.8
1.4
2.5
500 points
averaging
(± °C)
1.0
2.0
1.6
2.3
1.5
2.5
1.4
2.6
1.3
2.6
1.3
2.6
1.4
2.4
1.4
2.5
Table 2. Measurement accuracy of the U2355A at 0 to 18 °C and 28 to 45 °C
T/C type
B
E
J
K
N
R
S
T
Thermocouple measurement accuracy (U2355A at 0 to 18 °C and 28 to 45 °C)
ITS-90 Temperature range (°C)
Low
High
0
1820
â270
1000
â210
1200
â270
1372
â270
1300
â50
1768
â50
1768
â270
400
Optimum measurement range
(°C)
Low
High
1100
1820
400
1100
â150
1000
â200
â150
â150
1200
â210
â150
â100
1200
â200
â100
â100
1300
â200
â100
300
1760
â50
300
400
1760
â50
400
â100
400
â200
â100
21
Without
|averaging
(± °C)
3.4
7.5
2.7
3.8
2.5
4.2
2.9
4.3
2.6
4.9
3.8
8.5
4.2
7.7
2.4
4.3
50 points
averaging
(± °C)
2.4
3.6
2.6
3.6
2.4
4.0
2.8
4.0
2.5
4.2
3.1
4.6
3.4
4.2
2.2
4.3
500 points
averaging
(± °C)
2.2
2.2
2.5
3.6
2.4
3.9
2.8
3.9
2.5
4.0
3.0
3.3
3.2
3.1
2.2
3.9
|
▷ |