Accuracy specifications (continued)
Model
Bipolar Range (V)
10
5
2.5
1.25
1
0.5
0.25
0.2
0.05
Offset Error (mV)1
2.0
1.5
1.5
1.0
1.0
1.0
1.0
1.0
1.0
U2331A
Gain Error (mV)
6.0
4.0
2.0
1.5
1.0
1.0
1.0
1.0
1.0
–– The above specifications are typical for 23 °C.
–– Specifications are for 20 minutes warm-up and self calibration.
–– The measurement are calculated with 100 points averaging of data.
Accuracy (% of reading + offset error)2
0.06% + 2.0 mV
0.08% + 1.5 mV
0.08% + 1.5 mV
0.12% + 1.0 mV
0.10% + 1.0 mV
0.20% + 1.0 mV
0.40% + 1.0 mV
0.50% + 1.0 mV
2.02% + 1.0 mV
Model
Unipolar Range (V)
10
Bipolar Range (V)
10
Model
Unipolar Range (V)
10
Bipolar Range (V)
10
Model
Unipolar Range (V)
10
Bipolar Range (V)
10
Offset Error (mV)3
1.0
Offset Error (mV)3
1.0
Offset Error (mV)3
1.0
Offset Error (mV)3
1.0
Offset Error (mV)3
2.5
Offset Error (mV)3
1.5
U2351A, U2352A, U2353A, U2354A
Analog output
Gain Error (mV)
Accuracy (% of reading + offset error)4
2.0
0.02% + 1.0 mV
Gain Error (mV)
Accuracy (% of reading + offset error)4
4.0
0.04% + 1.0 mV
U2355A, U2356A
Gain Error (mV)
Accuracy (% of reading + offset error)4
2.0
0.02% + 1.0 mV
Gain Error (mV)
Accuracy (% of reading + offset error)4
4.0
0.04% + 1.0 mV
U2331A
Gain Error (mV)
4.0
Gain Error (mV)
4.0
Accuracy (% of reading + offset error)4
0.04% + 2.5 mV
Accuracy (% of reading + offset error)4
0.04% + 1.5 mV
–– The above specifications are typical for 23 °C.
–– Specifications are for 20 minutes warm-up and self calibration.
1 Offset error is measured at midscale of full range.
2 Accuracy = ± [% of |(Gain error / (Measured value – Midscale of FSR))| + Offset error]
3 Offset error is measured at 0 V.
4 Accuracy = ± [% of |Gain error/Output value| + Offset voltage]