English
Language : 

E1962B Datasheet, PDF (19/27 Pages) Agilent(Hewlett-Packard) – cdma2000/IS-95/AMPS Mobile Test Application
19 | Keysight | E1962B cdma2000/IS-95/AMPS Mobile Test Application - Technical Overview
CDMA analyzer (continued)
Tuned channel power measurement
Input frequency ranges 411 to 420 MHz
450 to 484 MHz
824 to 934 MHz
1700 to 1980 MHz
Measurement method
Measures the total power in a 1.23 MHz
bandwidth centered on the active reverse
channel center frequency
Measurement data
capture period
0.3125 ms (very fast mode), 1.25 ms (fast
mode), or 10 ms (normal mode)
Measurement trigger 20 ms clock (frame trigger)
Maximum input level +37 dBm/1.23 MHz peak (5 W peak)
Measurement range
–61 to +30 dBm, usable to < –69 dBm/
1.23 MHz with reduced accuracy
Measurement level
ranging
Auto and manual
Measurement accuracy < ± 1 dB 15 to 55 °C, typically < ± 0.5 dB, for the
normal and fast modes
< ± 1.1 dB 15 to 55 °C, typically < ± 0.5 dB, for
the very fast mode (calibrated against average
power and within ± 10 degrees of calibration
temperature; calibration must occur between
20 to 55 °C)
Measurement resolution 0.01 dBm/1.23 MHz
Measurement result Channel power in a 1.23 MHz bandwidth
Concurrency capabilities Channel power measurements can be made
concurrently with all CDMA measurements that
support concurrency
Calibrate function
Calibrates the channel power measurement
over the entire operating frequency range of the
test set against the average power measure-
ment; no external cabling is required
Calibration time
Typically < 120 s
Access probe power measurement
Input frequency ranges 411 to 420 MHz
450 to 484 MHz
824 to 934 MHz
1700 to 1980 MHz
Measurement method
Measures the total power in a 1.23 MHz
bandwidth centered on the active reverse
channel center frequency
Measurement data
capture period
1.25 ms
Measurement trigger Amplitude rise only
Maximum input level +37 dBm/1.23 MHz peak (5 W peak)
Measurement range −54 to +30 dBm
Measurement level
ranging
Auto and manual
Measurement accuracy
< ± 1 dB 15 to 55 °C, typically < ± 0.5 dB (cali-
brated against average power and within ± 10
degrees of calibration temperature; calibration
must occur between 20 to 55 °C)
Measurement result Access probe power in a 1.23 MHz bandwidth
Concurrency capabilities None
CDMA analyzer (continued)
Handoff modulation quality measurement
Input frequency ranges 411 to 484 MHz
800 to 1000 MHz
1700 to 2000 MHz
Measurement chip rate 1.2288 Mcps
Modulation
measurement method
PN offset handoff for IS-2000 R-RC3, R-RC4,
or R-RC8 to generate reverse pilot only pream-
ble; measures single code rho on the preamble
with HPSK (R-Pilot only); performs two hand-
offs: one to initiate the preamble and a second
to return the initial PN offset
Maximum input level +37 dBm/1.23 MHz peak (5 W peak)
Input level range
−25 to +30 dBm/1.23 MHz, usable to
−50 dBm/1.23 MHz with reduced accuracy
Modulation quality
measurement range (for
signals with < ± 6 μs
time error and < ± 1 kHz
frequency error)
0.40 to 1.00 rho
Measurement interval 1.042 ms (5 Walsh symbols)
Modulation quality mea- < ± 0.003 + residual error for 0.8 < rho < 1.0
surement accuracy
Modulation quality mea- Residual rho: > 0.999
surement residuals
Residual EVM: < 4% rms
Residual time error: ± 0.11 μs
Frequency error
± 15 Hz plus timebase error
Measurement results
rho, frequency error, time error, carrier feed-
through, phase error, amplitude error, and EVM
Concurrency capabilities None
Modulation quality measurement
Input frequency ranges 411 to 484 MHz
800 to 1000 MHz
1700 to 2000 MHz
Measurement chip rate 1.2288 Mcps
Modulation measure-
ment method
IS-95, R-RC1, or R-RC2: Single code rho
R-RC3, R-RC4, or R-RC8: Multi-code rho and
EVM with code domain results
Maximum input level +37 dBm/1.23 MHz peak (5 W peak)
Input level range
−25 to +30 dBm/1.23 MHz, usable to
−50 dBm/1.23 MHz with reduced accuracy
Modulation quality
measurement range (for
signals with < ± 6 μs
time error and < ± 1 kHz
frequency error)
IS-95, R-RC1, or R-RC2: 0.40 to 1.00 rho
R-RC3, R-RC4, or R-RC8: 1 to 40% EVM
Measurement interval 1.042 ms (5 Walsh symbols)
Modulation quality
IS-95, RC1, or RC2: < ± 0.003 + residual error
measurement accuracy for 0.8 < rho < 1.0
R-RC3, R-RC4, or R-RC8: < 1.25% rms +
residual error for 1% < EVM < 20%
Modulation quality
Residual rho: > 0.999
measurement residuals Residual EVM: < 4% rms, typically < 3.1% rms
Residual time error: ± 0.11 μs
Frequency error: ± 15 Hz plus timebase error
Residual code domain power: < −35 dBc