10 | Keysight | E1962B cdma2000/IS-95/AMPS Mobile Test Application - Technical Overview
Audio analyzer specifications
–– All specifications for the audio analyzer apply to signals pres-
ent at test set’s AUDIO IN ports.
–– Audio analyzer de-emphasis: 750 μs, de-emphasis settable
as off or on
–– Audio analyzer expander: Settable as off or on with reference
level setting of 10 mV to 10 V
–– Audio analyzer filters: Settable choices of none, C-message,
50 Hz to 15 kHz band pass, 300 Hz to 15 kHz band pass, or
100 Hz bandwidth tunable band pass tunable over 300 Hz to
15 kHz
Audio level measurement
Types of signals
measured
Sinusoidal audio signals
Measurement frequency 100 Hz to 15 kHz
range
AUDIO IN level range 7.1 mV to 20 V peak (5 mV to 14.1 V rms)
Measurement accuracy < ± (2% of reading + resolution) for 100 Hz to
8 kHz, < ± (3% of reading + resolution) for 8 to
15 kHz
Measurement THD plus < 200 μV rms
noise
Measurement detector Selectable choices of rms and peak
Measurement trigger
source
Immediate
Available result
Audio level
Multi-measurement
capabilities
1 to 999 measurements, average, minimum,
maximum, and standard deviation results
Concurrency capabilities Audio level measurements can be made
concurrently with all other measurements
External input
impedance
Typically 100 k W in parallel with 105 pF
Measurement resolution Typically 0.3% of expected level setting or
0.2 mV, whichever is greater
SINAD measurement
Types of signals
measured
Sinusoidal audio signals
Measurement frequency 100 Hz to 10 kHz
range
AUDIO IN level range 42.4 mV to 20 V peak (30 mV to 14.1 V rms)
Measurement accuracy < ± 1.0 dB for SINAD < 44 dB
Residual THD plus noise < –60 dB or 200 μV rms, whichever is greater
Measurement trigger
source
Immediate
Available result
SINAD ratio
Multi-measurement
capabilities
1 to 999 measurements, average, minimum,
maximum, and standard deviation results
Concurrency capabilities SINAD measurements can be made concurrent-
ly with all analog and audio measurements
Measurement resolution Typically 0.01 dB
Distortion measurement
Types of signals
measured
Sinusoidal audio signals
Measurement frequency 100 Hz to 10 kHz
range
AUDIO IN level range 42.4 mV to 20 V peak (30 mV to 14.1 V rms)
Measurement accuracy < ± 12% of reading (± 1.0 dB) for distortion >
0.67%
Residual THD plus noise < –60 dB or 200 μV rms, whichever is greater
Measurement trigger
source
Immediate
Available result
Audio distortion
Multi-measurement
capabilities
1 to 999 measurements, average, minimum,
maximum, and standard deviation results
Concurrency capabilities Distortion measurements can be made concur-
rently with all analog and audio measurements
Measurement resolution Typically 0.1%
Audio frequency measurement
Types of signals
measured
Sinusoidal audio signals
Measurement frequency 100 Hz to 15 kHz
range
AUDIO IN level range 7.1 mV to 20 V peak (5 mV to 14.1 V rms)
AUDIO IN signal
conditions
Signal at test set’s AUDIO IN must have
signal-to-noise ratio > 30 dB
Measurement accuracy < 0.1 Hz averaged over 10 measurements,
< 1.0 Hz for a single measurement
Measurement THD plus < 200 μV rms
noise
Measurement trigger
source
Immediate
Available result
Audio frequency
Multi-measurement
capabilities
1 to 999 measurements, average, minimum,
maximum, and standard deviation results
Concurrency capabilities Frequency measurements can be made concur-
rently with all other measurements
Measurement resolution Typically 0.1 Hz