U1210 Datasheet, PDF (14/17 Pages) Agilent(Hewlett-Packard) – Agilent U1210 Series Handheld Clamp Meters
14 | Keysight | U1210 Series Handheld Clamp Meters – Data Sheet
Electrical Specifications for U1213A (continued)
AC current
Accuracy *N1
Range
Resolution
45 ~ 65 Hz 65 ~ 200 Hz 200 ~ 300 Hz 300 ~ 400 Hz Maximum overload
40 A
0.01 A
2.0 % + 10 3.0 % + 10 3.5 % + 10 6.5 % + 10
400 A
0.1 A
2.0 % + 5
3.0 % + 5
3.5 % + 5
6.5 % + 5
1000 A R.M.S.
1000 A
1A
2.0 % + 5
3.0 % + 5
3.5 % + 5
6.5 % + 5
–– N1: The maximum verification of current and frequency product is less than 400,000 A x Hz
AC+DC current
Range
40 A
400 A
1000 A
Resolution
0.01 A
0.1 A
1A
Accuracy *N1
45 ~ 65 Hz
3.5 % + 25
3.5 % + 9
4.5 % + 9
65 ~ 200 Hz
4.5 % + 25
4.5 % + 9
5.0 % + 9
200 ~ 300 Hz
5.0 % + 25
5.0 % + 9
5.5 % + 9
300 ~ 400 Hz
8.0 % + 25
8.0 % + 9
8.5 % + 9
Maximum overload
1000 A R.M.S.
Current (1 ms peak hold)
Range
Resolution
40 A
0.01 A
400 A
0.1 A
1000 A
1A
–– Specified accuracy for changes > 1 ms in duration
Accuracy
2.0 % + 70
2.0 % + 43
2.0 % + 43
Maximum overload
1000 A R.M.S.
Temperature test
Thermal type
K
Range
–200 ~ –40 °C
–40 ~ 1372 °C
–328 ~ –40 °F
–40 ~ 2502 °F
Resolution
0.1 °C
0.1 °C
0.1 °F
0.1 °F
Accuracy
1 % + 3 °C
1 % + 1 °C
1 % + 6 °F
1 % + 2 °F
Notes:
1.  The accuracy does not include the tolerance of the thermocouple probe, and the meter must be operational for at least one hour
2.  Do not allow the temperature sensor to contact a surface that is energized above 33 V R.M.S. or 70 V DC. Such voltages will pose a shock
hazard.
3.  The temperature calculation is according to the standard of EN/IEC-60548-1 and NIST175