U1210 Datasheet, PDF (13/17 Pages) Agilent(Hewlett-Packard) – Agilent U1210 Series Handheld Clamp Meters | |||
|
◁ |
13 | Keysight | U1210 Series Handheld Clamp Meters â Data Sheet
Electrical Specifications for U1213A (continued)
Diode check/audible continuity test
Range
Resolution
Accuracy
Test current
Open voltage
Diode
0.001 V
0.5 % + 2
Approximately 0.8 mA
< +3.1 V
ââ Overload protection: 1000 V R.M.S. for circuits < 0.3 A of short circuit current
ââ Built-in buzzer sounds when reading is below 50 mV approximately and single tone for normal forward-biased diode
or semiconductor junction as 0.3 V ⤠Reading ⤠0.8 V
Resistance
Range
400 â¦
4 kâ¦
40 kâ¦
400 kâ¦
4 Mâ¦
40 Mâ¦
Resolution
0.1 â¦
0.001 kâ¦
0.01 kâ¦
0.1 kâ¦
0.001 Mâ¦
0.01 Mâ¦
Accuracy
0.3 % + 3
0.6 % + 3
2.0 % + 5
Test current
0.8 mA
80 µA
8 µA
727 nA
112 nA
112 nA
Notes:
1.â Overload protection: 1000 V R.M.S. for circuits < 0.3 A of short circuit current
2.â Maximum open voltage: < +3.1 V
3.â Instant continuity: Built-in buzzer sounds when resistance is less than 10.0 â¦
4.â The accuracy of 400 ⦠and 4 k⦠is specified after Relative function, which is used to substrate the test lead resistance and the thermal effect
Capacitance
Range
Resolution
Accuracy
Overload protection
4 µF
40 µF
400 µF
0.001 µF
0.01 µF
0.1 µF
1%+4
2%+4
1000 V R.M.S. for circuits
< 0.3 A of short circuit current
4000 µF
1 µF
3%+4
ââ The accuracy is based on film capacitor or better and use Relative mode to zero residual value
DC current
Range
Resolution
40 A
0.01 A
400 A
0.1 A
1000 A
1A
ââ Use Relative mode to zero residual offset
Accuracy
1.5 % + 15
1.5 % + 3
2.0 % + 5
Maximum overload
1000 A R.M.S.
|
▷ |