U1210 Datasheet, PDF (6/17 Pages) Agilent(Hewlett-Packard) – Agilent U1210 Series Handheld Clamp Meters
Electrical Specifications for U1211A
Accuracy is given as ± (% of reading + number of least significant digits) at 23 °C ± 5 °C, with relative humidity less than 80 % R.H.
ACV/ ACA specifications are ac coupled, true R.M.S. and are valid from 5 % of range to 100 % of range. The crest factor may be up to
3.0 at full-scale except the 1000 V and 1000 A ranges where these are 1.5 at full scale. For non-sinusoidal waveforms, add (2 % reading +
2 % full scale) typical, for crest factors up to 3.
DC voltage
Range
Resolution
400 V
0.1 V
1000 V
1V
• Input impedance: 10 MΩ (nominal)
Accuracy
0.5 % + 5
0.5 % + 3
Overload protection
1000 V R.M.S
AC voltage
Range
Resolution
400 V
0.1 V
1000 V
1V
• Input impedance: 10 MΩ (nominal) in parallel with < 100 pF
Accuracy
45~400 Hz
1%+5
1%+5
Overload protection
1000 V R.M.S
Voltage (1 ms peak hold)
Range
Resolution
400 V
0.1 V
1000 V
1V
• Specified accuracy for changes > 1 ms in duration
Accuracy
1 % + 43
1 % + 43
Overload protection
1000 V R.M.S
Diode check/audible continuity test
Range
Resolution
Accuracy
Test current
Open voltage
Diode
0.001 V
0.5 % + 2
Approximately 0.8 mA
< +3.1 V
• Overload protection: 1000 V R.M.S. for circuits < 0.3 A of short circuit current
• Built-in buzzer sounds when reading is below 50 mV approximately and single tone for normal forward-biased diode or semiconductor junction as
0.3 V ≤ Reading ≤ 0.8 V
Resistance
Range
400 Ω
4 kΩ
Resolution
0.1 Ω
0.001 kΩ
Accuracy
0.5 % + 3
0.5 % + 3
Test current
0.8 mA
80 µA
Notes:
1. Overload protection: 1000 V R.M.S. for circuits < 0.3 A of short circuit current
2. Maximum open voltage: < +3.1 V
3. Instant continuity: Built-in buzzer sounds when resistance is less than 10.0 Ω
4. The accuracy of 400 Ω and 4 kΩ is specified after Relative function, which is used to substrate the test lead resistance and the thermal effect