English
Language : 

U1210 Datasheet, PDF (5/17 Pages) Agilent(Hewlett-Packard) – Agilent U1210 Series Handheld Clamp Meters
General Specifications
Title
Dimension
Net Weight
Display
Battery
Low battery indicator
Power consumption
Battery life
Connectivity
Maximum jaw opening
Temperature coefficient
NMRR (Normal Mode Rejection Ratio)
CMRR (Common Mode Rejection Ratio)
Operating temperature
Storage temperature
Relative Humidity (R.H.)
Temperature coefficient
Safety compliance
Measurement category
EMC compliance
Specification
• U1211A: 106 mm (W) X 273 mm (L) X 43 mm (H)
• U1212A and U1213A: 106 mm (W) X 260 mm (L) X 43 mm (H)
• U1211A: 625 grams with batteries included
• U1212A and U1213A: 525 grams with batteries included
4 digits with maximum reading 4,100 counts. The 12 Segments analog bar graph and full
annunciator. Automatic polarity indication.
Standard 9 V Battery - Alkaline
Battery voltage drops below 6.0 V
• U1211A: 186 mVA maximum
• U1212A and U1213A: 220 mVA maximum
60 hours (typical)
Wireless Bluetooth compatible
~2 inches
0.12 x (specified accuracy)/°C (from 0 to 18 °C or 28 to 50 °C)
This series has a NMRR specification of > 60 dB at 50 Hz and 60 Hz, which means a good
ability to reject the effect of AC noise in DC measurement
• U1211A and U1212A have CMRR specifications of > 60 dB at DC to 60 Hz in the ACV
function; and > 80 dB at DC, 50 Hz and 60 Hz in the DCV function.
• U1213A has a CMRR specifications of > 60 dB at DC to 60 Hz in the ACV function;
and > 120 dB at DC, 50 Hz and 60 Hz in the DCV function.
–10 to 50 °C, 0 to 80 % R.H.
–20 to 60 °C, 0 to 80 % R.H.
Maximum 80 % R.H. for temperature up to 31 °C decreasing linearly to 50 % R.H. at 50 °C
0.1 × (specified accuracy) / °C (from 0 °C to 18 °C or 28 °C to 50 °C)
EN/IEC 61010-1:2001, ANSI/UL 61010-1:2004, and CAN/CSA-C22.2 No.61010-1-04
CAT III 1000 V/ CAT IV 600 V
• Certified to IEC61326-1:2005/ EN61326-1:2006
• CISPR 11:2003/ EN 55011:2007 Group 1 class A
• Canada: ICES-001:2004
• Australia/New Zealand: AS/NZS CISPR11:2004
4