Electrical Specifications for U1213A (continued)
AC current
Accuracy *N1
Range
Resolution
45~65 Hz
65 Hz~1 KHz
40 A
0.01 A
2.0 % + 10
3.0 % + 10
400 A
0.1 A
2.0 % + 5
3.0 % + 5
1000 A
1A
2.0 % + 5
3.0 % + 5
• N1: The maximum verification of current and frequency product is less than 400,000 A x Hz
Maximum overload
1000 A R.M.S.
AC+DC current
Range
40 A
400 A
1000 A
Resolution
0.01 A
0.1 A
1A
Accuracy
45~65 Hz
3.5 % + 25
3.5 % + 9
4.5 % + 9
65 Hz~1 kHz
4.5 % + 25
4.5 % + 9
5.0 % + 9
Maximum overload
1000 A R.M.S.
Current (1 ms peak hold)
Range
Resolution
40 A
0.01 A
400 A
0.1 A
1000 A
1A
• Specified accuracy for changes > 1 ms in duration
Accuracy
2.0 % + 70
2.0 % + 43
2.0 % + 43
Maximum overload
1000 A R.M.S.
Temperature test
Thermal type
K
Range
–200 ~ –40 °C
–40 ~ 1372 °C
–328 ~ –40 °F
–40 ~ 2502 °F
Resolution
0.1 °C
0.1 °C
0.1 °F
0.1 °F
Accuracy
1 % + 3 °C
1 % + 1 °C
1 % + 6 °F
1 % + 2 °F
Notes:
1. The accuracy does not include the tolerance of the thermocouple probe, and the meter must be operational for at least one hour
2. Do not allow the temperature sensor to contact a surface that is energized above 33 V R.M.S. or 70 V DC. Such voltages will pose a shock hazard.
3. The temperature calculation is according to the standard of EN/IEC-60548-1 and NIST175