Fast Device Tune Measurement
Measurement method
Allows simultaneous calibration
of a deviceâs transmitter output
power and receiver input level
across level and frequency in a
single sweep (per frequency
band). The device must operate
in a test mode, which forces it
to transmit a predeï¬ned series
of power steps at various
reverse link frequencies, and
simultaneously tune its receiver
to perform measurements (such
as RSSI) of the test setâs signal
at various forward link
frequencies and power levels
Measurement range
â65 dBm/1.23 MHz to
+28 dBm/1.23 MHz
Measurement interval
1.25 ms
Measurement ï¬lter
1.23 MHz brick wall
Step size range
5 to 20 ms
Trigger sources
Immediate and RF rise
Output level setting (to be within 0.1 dB of the ï¬nal power level)
Amplitude switch
< 1 ms, typically < 250 µs
Frequency switch
< 1 ms, typically < 500 µs
within band
Measurement accuracy1 (within 45 dB of expected power with
< 48 hours warm-up before self alignment initiation)
â54 to +28 dBm/1.23 MHz < ±0.75 dB, typically < ±0.3 dB
â61 to < â55 dBm/1.23 MHz < ±0.85 dB, typically < ±0.3 dB
RF source output level range
RF in/out port
â115 dBm/1.23 MHz to
â15 dBm/1.23 MHz
RF out only port
â115 dBm/1.23 MHz to
â5 dBm/1.23 MHz
Composite absolute output level accuracy (< ±10 °C and
< 24 hours from the last alignment, â108 to â15 dBm/1.23 MHz)
RF in/out port
< ±1.0 dB, typically ±0.65 dB
RF out only port
< ±1.0 dB, typically ±0.65 dB
RF output level settling
< 1 ms to be within 0.1 dB of the
ï¬nal value
1. Additional accuracy error when using RF OUT ONLY port is < ±0.1 dB.