E6890A Datasheet, PDF (4/6 Pages) Agilent(Hewlett-Packard) – Wireless Communications Test Set | |||
|
◁ |
RF Channel Suite â continued
Spectrum Monitor
Power versus time measurement
This measurement is a graphical view of output power in the time
domain. It is also useful as a zero-span spectrum analyzer.
Dynamic range (â15 dBm input signal)
With 1, 30, 100, 300, 640 kHz Typically > 64 dB
GSM Tx power ï¬lters
With 1.23, 3.84, 5.0 MHz
Typically > 54 dB
W-CDMA mean power ï¬lters
IQ capture measurement
This measurement returns the IQ samples collected during the
most recent RF channel suite measurements in either rectangular
or polar format. The collected samples are provided as real/imagi-
nary number pairs in rectangular format, or as magnitude/phase
pairs in polar format.
Input level range
Average power
â65 to +35 dBm
Absolute level accuracy¹
Within cellular frequency bands at expected frequency and level
â59 to +35 dBm
â65 to < â59 dBm
< ±0.6 dB, typically < ±0.3 dB
< ±0.7 dB, typically < ±0.4 dB
Within cellular frequency bands at expected frequency and level
with < 48 hours warm-up before self alignment initiated
â59 to +35 dBm
â65 to < â59 dBm
< ±0.7 dB, typically < ±0.3 dB
< ±0.8 dB, typically < ±0.4 dB
380 to 2000 MHz at expected frequency and level
â59 to +35 dBm
â65 to < â59 dBm
Typically < ±0.4 dB
Typically < ±0.6 dB
Measurement results
IQ samples
Number of samples
Sample period
QPSK EVM measurement
This measurement calculates QPSK composite EVM and several
other results relating to UE modulation quality.
Input level range
Average power
â25 to +28 dBm/3.48 MHz
UE ranges
EVM
Frequency error
⤠25% rms
< ±10 kHz
Measurement accuracy (Includes residual results,
measured for one slot (0.66666667 ms) measurement interval)
UE EVM 0% rms
< 2.5% rms
UE EVM 17.5% rms
< 0.7% rms
UE 25% rms
< 0.5% rms
Frequency error
< (±10 Hz + timebase accuracy)
Filter
5 MHz
Measurement results
EVM, phase error, magnitude
error, frequency error, origin
offset
Absolute level ï¬atness
Typically < ±0.15 dB
Spurious response
Typically < â65 dBc
Signals ⥠â10 dBm excluding spurs that change with input
attenuation but not with input level
Resolution bandwidth (RBW)
Span
Amplitude scaling
Markers
Number
Frequency
Amplitude
3 kHz to 5 MHz
RBW to 25 MHz
0.1 to 20 dB/division
0 to 5
Absolute, relative
Absolute, relative
1. Additional accuracy error when using RF OUT ONLY port is < ±0.1 dB.
4
|
▷ |