E6890A Datasheet, PDF (2/6 Pages) Agilent(Hewlett-Packard) – Wireless Communications Test Set | |||
|
◁ |
Technical Speciï¬cations
Amplitude modulation
These speciï¬cations apply to an E6601A mainframe and the
E6890A General Purpose Application ï¬rmware revision A.04 or
higher. Speciï¬cations describe the test setâs warranted perfor-
mance and are valid for the unitâs operation within ±10 °C of the
last self alignment. All speciï¬cations are valid after a 30-minute
warm-up period of continuous operation with valid self-alignment
unless otherwise noted. If the instrument has been off for longer
than 48 hours, a 48-hour warm-up period followed by self-align-
ment is required.
Modulation frequency range
AM depth range
AM accuracy
Total harmonic distortion
100 Hz to 100 kHz
0 to 99.9%
< ± 1.0% (20 kHz modulation
frequency, 60% AM depth and
â25 dBm output level)
< 0.5% (20 kHz modulation
frequency, 60% AM depth and
â25 dBm output level)
Supplemental characteristics are intended to provide typical, but
non-warranted, performance parameters that may be useful in
applying the instrument. These characteristics are shown in italics
and labeled as âtypical.â All units shipped from the factory meet
these typical numbers at +25 °C ambient temperature without
including measurement uncertainty.
CW RF Generator
Frequency
Frequency range
Accuracy
380 to 2700 MHz
Same as timebase accuracy
CW output level
Output level ranges
RF IN/OUT
â130 to â13 dBm
typical over-range to â10 dBm
RF OUT ONLY
â120 to â3 dBm
typical over-range to 0 dBm
Absolute level accuracy
(⤠±10 °C and ⤠24 hours from last self alignment)
â108 to â13 dBm
< ±1.0 dB,
at RF IN/OUT
typically < ±0.5 dBm
â108 to â5 dBm
at RF OUT ONLY
< ±1.0 dB,
typically < ±0.5 dBm
Setting resolution
0.01 dB
Maximum applied reverse power
RF IN/OUT
< +37 dBm (5 W) peak
RF OUT ONLY
< +24 dBm (0.25 W) peak
Frequency modulation
Modulation frequency range 10 Hz to 100 kHz
FM deviation range
0 to 100 kHz
FM deviation accuracy
< 3.5%
Total harmonic distortion
< 0.5%
Residual FM (0.3 - 3 kHz bandwith)
RF frequencies
< 1 GHz to 5 Hz rms
RF frequencies
2 to 2.2 GHz to 7 Hz rms
RF frequencies
> 2.2 GHz to 9 Hz rms
VSWR
RF IN/OUT
380 to 1000 MHz
1000 to 2000 MHz
2000 to 2200 MHz
2200 to 2700 MHz
< 1.15:1
< 1.2:1
< 1.3:1
< 1.35:1
RF OUT ONLY
380 to 1000 MHz
< 1.3:1
1000 to 2700 MHz< 1.5:1
Spectral purity
Harmonics
RF IN/OUT
â130 to â15 dBm,
400 to 2200 MHz
< â30 dBc
RF OUT ONLY
â120 to â5 dBm,
400 to 2200 MHz
< â30 dBc
Non-harmonic spurious
(< ±10 °C from last self alignment, 400 to 2200 MHz)
RF IN/OUT and RF OUT ONLY < â40 dBc, typically < â45 dBc
2
|
▷ |