E6890A Datasheet, PDF (3/6 Pages) Agilent(Hewlett-Packard) – Wireless Communications Test Set | |||
|
◁ |
RF Analyzer
Frequency ranges
Cellular bands
General purpose
Input level ranges
Average power
Peak power
Self alignment validity
VSWR
RF IN/OUT
400 to 1000 MHz
1000 to 2000 MHz
2000 to 2200 MHz
2200 to 2600 MHz
Filters
Trigger setup
Arm
Delay (varies by ï¬lter
and measurement)
Sources
(varies by measurement)
Measurement setup
Averaging
(multi-measurement
countânot applicable to
all measurements)
Timeout
411 to 486 MHz
776 to 960 MHz
1574 to 1577 MHz
1710 to 1980 MHz
400 to 2000 MHz
â65 to +33 dBm, typical
over-range to +35 dBm
â65 to +37 dBm (5 W)
⤠±10 °C change and ⤠30 days
from last self alignment
< 1.15:1
< 1.2:1
< 1.3:1
< 1.35:1
1 kHz
30 kHz
100 kHz
300 kHz
640 kHz
1.23 MHz
1.6 MHz
3.84 MHz
5.0 MHz
GSM Tx power
W-CDMA mean power
Single, continuous
â180 to 180 ms
External, fall, immediate, rise
Off, 1 to 999
Off, 0.1 to 999.9 s
RF Channel Suite
Includes channel power, frequency error, power versus time, and
IQ capture measurements.
Input level range
Average power
â65 to +35 dBm
Measurement interval and ï¬lter ranges
1 kHz ï¬lter
1 to 5000 ms
30 kHz ï¬lter
0.4 to 4000 ms
100 kHz ï¬lter
0.1 to 4000 ms
300 kHz ï¬lter
0.01 to 1700 ms
640 kHz ï¬lter
0.01 to 810 ms
1.23 MHz ï¬lter
0.01 to 420 ms
1.6 MHz ï¬lter
0.01 to 320 ms
3.84 MHz ï¬lter
0.01 to 135 ms
5.0 MHz ï¬lter
0.01 to 100 ms
GSM Tx power ï¬lter
0.01 to 199 ms
W-CDMA mean power ï¬lter 0.1 to 3200 ms
Channel power measurement
Measurement accuracy¹
Within cellular frequency bands
â59 to +35 dBm
< ±0.6 dB, typically < ±0.3 dB
â65 to < â59 dBm
< ±0.7 dB, typically < ±0.4 dB
Within cellular frequency bands with < 48 hours warm-up before
self alignment initiated
â59 to +35 dBm
< ±0.7 dB, typically < ±0.3 dB
â65 to < â59 dBm
< ±0.8 dB, typically < ±0.4 dB
400 to 2000 MHz
â59 to +35 dBm
â65 to < â59 dBm
Typically < ±0.4 dB
Typically < ±0.6 dB
Measurement repeatability Typically < ±0.05 dB
Returning to same level and frequency, no temperature change
and insigniï¬cant time change
Frequency error measurement
Measurement accuracy
CW signals from 400 to 2000 MHz
â60 to +35 dBm
with 1, 30, 100,
300, 640 kHz GSM
Tx power ï¬lters
< ±(50 Hz + timebase accuracy)
â40 to +35 dBm
with 1.23, 3.84,
5.0 MHz W-CDMA
mean power ï¬lters
< ±(50 Hz + timebase accuracy)
1. Additional accuracy error when using RF OUT ONLY port is < ±0.1 dB.