English
Language : 

S11155-2048-01_15 Datasheet, PDF (6/8 Pages) Hamamatsu Corporation – CCD linear image sensors
CCD linear image sensors
S11155-2048-01, S11156-2048-01
ARG
REGH, REGL
TG
P1H
P2H
SG
RG
OS
MPP operation
Tpwv Tovr
1 line output period
Tpwar
(electronic shutter: closed)
Tinteg
(electronic shutter: open)
(REGH=-4 V, REGL=-6.5 V)
Tpwreg
(REGH, REGL=-8 V)
Tpwh, Tpws
1
2
3..2067
2068 2069
2070...
N*
Tpwr
D1
D2
D19 D20
D3..D10, S1...S2048, D11..D18
Normal readout period
Dummy readout period
* Apply clock pulses to the specified terminals during the period of dummy readout.
Set the total number of clock pulses N, according to the integration time.
Parameter
Symbol
Min.
Typ.
Pulse width
Tpwar
*21
-
ARG
Rise and fall times
Tprar, Tpfar
200
-
REGH, REGL
Pulse width
Rise and fall times
Tpwreg
-
Tprreg, Tpfreg
100
Tinteg - Tpwv
-
TG
Pulse width
Tpwv
2
Rise and fall times
Tprv, Tpfv
20
-
-
Pulse width
Tpwh
50
100
P1H, P2H*22
Rise and fall times
Tprh, Tpfh
10
-
Duty ratio
-
40
50
Pulse width
Tpws
50
100
SG
Rise and fall times
Tprs, Tpfs
10
-
Duty ratio
-
40
50
RG
Pulse width
Rise and fall times
Tpwr
Tprr, Tpfr
5
5
15
-
TG - P1H
Overlap time
Tovr
1
2
Integration time
Tinteg
2
-
*21: The Min. value of Tpwar is equal to the normal readout period.
*22: Symmetrical clock pulses should be overlapped at 50% of maximum pulse amplitude.
Max.
-
-
-
-
-
-
-
-
60
-
-
60
-
-
-
-
KMPDKCMP0DC3034477EDE
Unit
μs
ns
μs
ns
μs
ns
ns
ns
%
ns
ns
%
ns
ns
μs
μs
6