English
Language : 

THS3000 Datasheet, PDF (2/12 Pages) List of Unclassifed Manufacturers – Handheld Oscilloscopes
Datasheet
Four isolated input channels easily handle any type of mixed signal inputs
Safely make floating and differential
measurements
Input signal and float voltage maximum safety ratings
Selecting the right probe
Scope/Probe Maximum safety rating THS3000 viewable signal
(attenuation) Reference Input signal On-screen P- On-screen
float safety safety rating P voltage RMS voltage
rating
THS3000 (1X 600 VRMS
input)
CAT III
300 VRMS
CAT III
800 Vp-p
282 VRMS
1000 VRMS
CAT II
THP0301
(10X)
300 VRMS
CAT III
300 VRMS
CAT III
849 Vp-p
300 VRMS
PP5122
(100X)
600 VRMS
CAT II
1000 VRMS
CAT III
2828 Vp-p
1000 VRMS
P5122 (100X) 600 VRMS
CAT II
1000 VRMS
CAT III
2828 Vp-p
1000 VRMS
Isolated-channel technology is specified to 1000 VRMS maximum float voltage
Making accurate and safe measurements on power electronics, power
semiconductors, and other electronics applications can be challenging
when the signal reference is floating and not referenced to earth ground.
When your signal ranges from low voltage to high voltage (kV) or you must
use probing techniques that can potentially create ground loops, the
problem is compounded.
To enable floating measurements, the THS3000 Series is architecturally
different than most other oscilloscopes. All input channels are fully isolated
from the main chassis and from each other. Additionally the power adapter
and USB interface are fully isolated to ensure safe measurements and
eliminate the risk of unintentional grounding or accidental short circuits.
When configured with the proper probes you can be assured you'll be able
to make quick, safe, and accurate measurements.
Fast analysis of your device
The THS3000 Handheld Oscilloscope Series is packed with analysis tools
to help you validate your device's operation and identify issues quickly. With
waveform math, you can add, subtract, or multiply any of your signals to
investigate instantaneous power or look at gain. Vertical and horizontal
cursors allow you to look at a specific point on your waveform for accurate
measurements of voltage, current, time, or frequency. With 21 automated
measurements, you can make common measurements quickly and
accurately. The built-in Fast Fourier Transform (FFT) function allows you to
see the frequency spectrum of your signal, revealing signal interference,
crosstalk, or switching noise.
2 www.tektronix.com