English
Language : 

CNX3000 Datasheet, PDF (4/7 Pages) List of Unclassifed Manufacturers – Fluke CNX3000 Series Test Tools
Fluke CNX 3000 Series AC Voltage Module
Accuracy is specified as ± ([% of reading] + [number of least significant digits]). All ranges
are autoranging. Accuracy is specified from 5 % to 100 % of the range obtained by
autoranging, from 18 °C to 28 °C.
AC voltage
Range1
Resolution
Accuracy2,3
45 Hz to 500 Hz
500 Hz to 1 kHz
6.000 V
0.001 V
60.00 V
600.0 V
0.01 V
0.1 V
1.0 % + 3
2.0 % + 3
1000 V
1V
1 All ac voltage ranges are specified from 1 % of range to 100 % of range.
2 Crest factor of ≤ 3 at full scale up to 500 V, decreasing linearly to crest factor ≤ 1.5 at 1000 V.
3 For non-sinusoidial waveforms, add –(2 % of reading + 2 % full scale) typical, for crest factor up to 3.
General specifications
LCD w/backlight
Battery type
Battery life
Memory
RF communications
RF communication range
Operating temperature
Storage temperature
Temperature coefficient
Operating humidity
EMC
Safety compliance
Safety rating
Certifications
Ingress Protection (IP) rating
Size (HxWxD)
Weight
3½ digits, 6000 counts, updates 4/sec
2 AA, NEDA 15 A, IEC LR6
400 Hours
Record up to 65000 readings
2.4 GHZ ISM Band
20 meters
-10 °C to +50 °C
-40 °C to +60 °C
0.1 X (specified accuracy)/°C (< 18 °C or > 28 °C)
90 % at 35 °C, 45 % at 40 °C, 45 % at 50 °C
EN 61326-1:2006
EN/IEC 61010-1:2010 to 1000 V Measurement Category (CAT) III
600 V Measurement Category (CAT) IV
EN/IEC 61010-2-030:2010
EN/IEC 61010-031:2002+A1:2008
CAT IV 600 V, CAT III 1000 V
CSA, FCC T68-FWCS IC:6627A-FWCS
IP42
16.5 cm x 6.35 cm x 1.4 cm (6.5 in x 2.5 in x 1.4 in)
.22 kg (8 oz)
Contents
4 Fluke Corporation Fluke CNX 3000 Series Test Tools. The Fluke Wireless Team.