CNX3000 Datasheet, PDF (2/7 Pages) List of Unclassifed Manufacturers – Fluke CNX3000 Series Test Tools
Specifications
Fluke CNX 3000 Series Wireless Multimeter
Detailed specifications
For all specifications: accuracy is specified for one year after calibration, at operating tem-
peratures of 18 °C to 28 °C, with relative humidity at 0 % to 90 %. Accuracy specifications
take the form of ± ([ % of Reading ] + [ Number of least significant digits ]).
AC voltage
Accuracy2,3
Range1
Resolution
45 Hz to 500 Hz
500 Hz to 1 kHz
600.0 mV
0.1 mV
6.000 V
0.001 V
60.00 V
0.01 V
1.0 % + 3
2.0 % + 3
600.0 V
0.1 V
1000 V
1V
1 All ac voltage ranges are specified from 1 % of range to 100 % of range.
2 Crest factor of ≤ 3 at full scale up to 500 V, decreasing linearly to crest factor < 1.5 at 1000 V.
3 For non-sinusoidial waveforms, add –(2 % of reading + 2 % full scale) typical, for crest factor up to 3.
DC voltage, continuity, resistance, diode test and capacitance
Function
Range
Resolution
Accuracy
mV
600.0 mV
0.1 mV
0.09 % + 2
V
6.000 V
0.001 V
60.00 V
0.01 V
0.09 % + 2
600.0 V
0.1 V
1000 V
1V
0.15 % + 2
Ω
600 Ω
1Ω
Meter beeps at < 25 Ω, beeper detects opens or shorts
of 250 μs or longer.
Ω
600.0 Ω
0.1 Ω
0.5 % + 2
6.000 kΩ
0.001 kΩ
60.00 kΩ
60.00 kΩ
0.01 kΩ
0.1 kΩ
0.5 % + 1
600.0 kΩ
0.001 MΩ
50.00 MΩ
0.01 MΩ
1.5 % + 3
Diode test
2.000 V
0.001 V
1%+2
µF
1000 nF
1 nF
10.00 μF
0.01 μF
1.2 % + 2
100.0 μF
0.1 μF
9999 μF1
1 μF
10 % typical
1 In the 9999 μF range for measurements to 1000 μF, the measurement accuracy is 1.2 % + 2.
AC and DC current
Function
mA AC (45 Hz to 1 kHZ)
mA DC2
Range1
60.00 mA
400.0 mA3
60.00 mA
400.0 mA
Resolution
0.01 mA
0.1 mA
0.01 mA
0.1 mA
1 All ac current ranges are specified from 5 % of range to 100 % of range.
2 Input burden voltage (typical): 400 mA input 2 mV/mA.
3 400.0 mA accuracy specified up to 600 mA overload.
Accuracy
1.5 % + 3
0.5 % + 3