English
Language : 

MDS-040APS24B Datasheet, PDF (5/10 Pages) Delta Electronics, Inc. – MDS Medical AC-DC Open Frame
T``E`CHNICAL DATASHEET
MDS Medical AC-DC Open Frame
24Volt, 40Watt / MDS-040APS24 B
EMC (Compliant with IEC 60601-1-2 4th Ed. Requirements)
EMC / Emissions
EN55011/EN55022,FCC Title 47:Class B
Harmonic Current Emissions
Immunity to
Voltage Flicker
Electrostatic Discharge
Radiated Field
Electrical Fast Transient / Burst
Surge
Conducted
Power Frequency Magnetic Fields
Voltage Dips
IEC61000-3-2
IEC61000-3-3
IEC61000-4-2
IEC61000-4-3
IEC61000-4-4
IEC61000-4-5
IEC61000-4-6
IEC61000-4-8
IEC61000-4-11
Level 4 Criteria A1)
Air Discharge: 15kV
Contact Discharge: 8kV
Criteria A1)
80MHz-2700MHz, 10V/m AM modulation
385MHz-5785MHz, 28V/m Pulse mode and other
modulation
Level 3 Criteria A1) :2kV
Level 3 Criteria A1)
Common Mode3): 2kV
Differential Mode4): 1kV
Level 2 Criteria A1)
150kHz-80MHz, 3Vrms, 6Vrms at ISM bands and
Amateur radio bands
Criteria A1)
Magnetic field strength 30A/m
Criteria A1)
0% UT, 0.5 cycle (10ms)
, 0o/45o/90o/135o/180o/225o/270o/315o/360o
Criteria B2)
0% UT, 1 cycle (20ms), 0o
Criteria A1)
70% UT, 25 cycle (500ms), 0o
Criteria B2)
0% UT, 250 cycle (5000ms), 0o
1) Criteria A: Normal performance within the specification limits
2) Criteria B: Output out of regulation, or shuts down during test. Automatically restore to normal operation after test.
3) Asymmetrical: Common mode (Line to earth)
4) Symmetrical: Differential mode (Line to line)
5
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (November 2016, Rev. 02)